Skip to content

NA 389.829Recalls & reported problems

Synthes GmbH · 3.5MM HEXAGONAL SCREWDRIVER SHAFT-6MM HXC · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034776835), received between Aug 8, 2019 and Nov 29, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Material Integrity Problem1
  • Material Twisted/Bent1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 29, 2023Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT ON NOVEMBER 2, 2023, THE SCREWDRIVER STRIPPED WHILE USING THE FINAL TIGHTENER DURING A LATERAL CORPECTOMY PROCEDURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. ALTERNATIVE INSTRUMENTS IN THE SET WERE UTILIZED. THIS REPORT INVOLVES ONE 3.5MM HEXAGONAL SCREWDRIVER SHAFT-6MM HXC. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18226835 ↗
  • MalfunctionJun 3, 2021Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE WAS RECEIVED DAMAGED AT THE LOCAL DC. THE DEVICE WAS UNABLE TO ATTACH TO THE SCREWDRIVER SHAFT, AND COULDN¿T HOLD THE SCREW. THIS REPORT IS FOR A 3.5MM HEXAGONAL SCREWDRIVER SHAFT-6MM HXC. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11935794 ↗
  • MalfunctionAug 8, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING ANTERIOR LUMBAR INTERBODY FUSION (ALIF) L4-L5 PROCEDURE ON (B)(6) 2019, THE HEXAGONAL SCREWDRIVER FROM THE ANTERIOR TENSION BAND (ATB) PLATE SYSTEM WAS NOT LOCKING OFF AND HAS NO CLEAR CLICK SOUND WHEN LOCKING THE SCREWS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER ATB SET. THERE WAS NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.  UPON MANUFACTURER'S INVESTIGATION OF THE RECEIVED COMPLAINT DEVICE IT CAN BE SEEN THAT THE HEXAGONAL TIP IS SLIGHTLY TWISTED AS WELL SOME EDGES ARE DEFORMED CONCOMITANT DEVICE/S REPORTED: UNKNOWN SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN), SCREWDRIVER (PART # 287110550, LOT # UNKNOWN, QUANTITY 1)  THIS REPORT IS FOR ONE (1) 3.5MM HEXAGONAL SCREWDRIVER SHAFT-6MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8875562 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page