Skip to content

NA 489.174Recalls & reported problems

Synthes GmbH · 5.5MM TI CANCELLOUS LOCKING SCREW 54MM THRD LENGTH · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034784366), received between Jul 21, 2022 and Jul 21, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Integrity Problem2
  • No Apparent Adverse Event1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 21, 2022Material Integrity Problem; No Apparent Adverse Event; Material Integrity Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON PERFORMED AN L2 LATERAL CORPECTOMY WITH SUPPLEMENTAL PLATE FIXATION FROM L1-L3 WITH DEPUY SYNTHES TSLP LATERAL PLATE ON (B)(6) 2021. THE SURGEON GOT X-RAYS IN JUNE OF 2022 SHOWING THE EXPANDABLE CAGE HAD SUBSIDED INTO THE L1 VERTEBRAL BODY WITH ONE OF THE SCREWS FROM THE TLSP PLATE IN L1 BODY BROKEN AT THE NECK OF THE SCREW CLOSE TO THE SCREW/PLATE INTERFACE. THE PATIENT WAS STABLE WITH NO CLINICAL ISSUES. THE PATIENT HAD EXPERIENCED A POST-OP DEVICE MALFUNCTION THIS REPORT IS FOR ONE (1)5.5MM TI CANC LOCK SCRW 54MM THRD LGTH THIS IS REPORT 5 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15080333 ↗
  • InjuryJul 21, 2022Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON PERFORMED AN L2 LATERAL CORPECTOMY WITH SUPPLEMENTAL PLATE FIXATION FROM L1-L3 WITH DEPUY SYNTHES TSLP LATERAL PLATE ON (B)(6) 2021. THE SURGEON GOT X-RAYS IN JUNE OF 2022 SHOWING THE EXPANDABLE CAGE HAD SUBSIDED INTO THE L1 VERTEBRAL BODY WITH ONE OF THE SCREWS FROM THE TLSP PLATE IN L1 BODY BROKEN AT THE NECK OF THE SCREW CLOSE TO THE SCREW/PLATE INTERFACE. THE PATIENT WAS STABLE WITH NO CLINICAL ISSUES. THE PATIENT HAD EXPERIENCED A POST-OP DEVICE MALFUNCTION. THIS REPORT IS FOR ONE (1)5.5MM TI CANC LOCK SCRW 54MM THRD LGTH THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15080320 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page