Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034785592), received between Oct 26, 2021 and Oct 26, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Integrity Problem1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 26, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT, THE PATIENT UNDERWENT REVISION SURGERY DUE TO FRACTURE OF THE RODS BILATERALLY AT APPROXIMATELY T10 LEVEL (ABOVE CORPECTOMY SITE) AND COLLAPSE OF THE TI SYNMESH CAGE, WITH EVIDENCE THE CAGE FRACTURED/BROKE AND FELL INTO ITSELF. LATER THE PATIENT UNDERWENT A T10/11 CORPECTOMY WITH CAGE PLACEMENT AND T8-L1 PEDICLE SCREW FIXATION (B)(6) 2018. THE PATIENT PRESENTED WITH PAIN AROUND (B)(6) 2021 AND SUBSEQUENT IMAGING SHOWED THE FOLLOWING HARDWARE FAILURE: 1. FRACTURE OF THE RODS BILATERALLY AT APPROXIMATELY T10 LEVEL (ABOVE CORPECTOMY SITE) 2. COLLAPSE OF THE TI SYNMESH CAGE, WITH EVIDENCE OF THE CAGE FRACTURED/BROKE AND FELL INTO ITSELF. ON (B)(6) 2021, THE SURGEON REVISED THE BROKEN RODS POSTERIORLY AND PLACED FOUR COCR RODS WITH PARALLEL CONNECTORS TO STABILIZE THE SPINE. THE SURGEON RESECTED SCAR TISSUE POSTERIORLY TO FREE UP THE DAMAGED CAGE AS MUCH AS POSSIBLE. AFTER THE INTRA-OPERATIVE ASSESSMENT, THE SURGEON DETERMINED IT WOULD BE BEST TO REVISE THE CAGE ANTERIORLY WITH THE HELP OF THE THORACIC AND VASCULAR SURGEONS. ON (B)(6) 2021, THE SURGEON WAS ABLE TO SAFELY EXTRACT THE BROKEN CAGE. SCAR TISSUE HAD FORMED ACROSS THE T10/11 LEVEL, ADHERING THE CAGE TO V

    Patient problems: Pain

    MAUDE report 12697468 ↗