Click'X 498.990Recalls & reported problems
Synthes (U.S.A.) LP · 6.2MM TI CLICK'X® PEDICLE SCR PREASSEMBLED 50MM THRD LENGTH · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034792392), received between Jun 11, 2018 and May 7, 2019 — classified as 3 injuries.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Unintended Movement2
- Device Operates Differently Than Expected1
Patient problems
- No Code Available3
- Pain2
- Failure of Implant1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 7, 2019Unintended Movement; Unintended Movement
ON 06/18/2019: UPDATED EVENT DESCRIPTION: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ELECTIVE REVISION SURGERY SYSTEM INSTABILITY DUE TO THE CLICK COLUMN SYSTEM HAD LOOSENED, SPECIFICALLY, THE 3D HEADS, LOCKING SCREWS AND FOUR (4) PEDICLE SCREWS. THE PATIENT PRESENTED PAIN BEFORE DETECTING THE MALFUNCTION. THE CLICK COLUMN SYSTEM MALFUNCTION WAS DETECTED (B)(6) 2019. IT WAS MENTIONED THAT THE SCREWS THAT WERE USED IN THE SURGERY WERE PRE-ASSEMBLED FROM CLICK COLUMN SYSTEM THAT IS TO SAY THAT IN THIS SURGERY THEY DID NOT USE THE 3D HEADS. THE DEVICES WERE ORIGINALLY IMPLANTED ON (B)(6) 2019. THE RECOVERED IMPLANTS WERE, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 45, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 50, TWO (2) SOFT CURVED RODS AND FOUR (4) CLICKX LOCKING CAPS. THE SURGEON WILL MONITOR THE EFFECTIVENESS OF THE CLICK COLUMN SYSTEM THROUGH POSTOPERATIVE RADIOGRAPHIC CONTROLS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CURRENTLY, THE PATIENT WAS IN GOOD CONDITION. THIS COMPLAINT INVOLVES TEN (10) DEVICES.
Patient problems: No Code Available; Pain
MAUDE report 8587641 ↗InjuryMay 7, 2019Unintended Movement; Unintended Movement
ON 06/18/2019: UPDATED EVENT DESCRIPTION: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ELECTIVE REVISION SURGERY SYSTEM INSTABILITY DUE TO THE CLICK COLUMN SYSTEM HAD LOOSENED, SPECIFICALLY, THE 3D HEADS, LOCKING SCREWS AND FOUR (4) PEDICLE SCREWS. THE PATIENT PRESENTED PAIN BEFORE DETECTING THE MALFUNCTION. THE CLICK COLUMN SYSTEM MALFUNCTION WAS DETECTED (B)(6) 2019. IT WAS MENTIONED THAT THE SCREWS THAT WERE USED IN THE SURGERY WERE PRE-ASSEMBLED FROM CLICK COLUMN SYSTEM THAT IS TO SAY THAT IN THIS SURGERY THEY DID NOT USE THE 3D HEADS. THE DEVICES WERE ORIGINALLY IMPLANTED ON (B)(6) 2019. THE RECOVERED IMPLANTS WERE, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 45, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 50, TWO (2) SOFT CURVED RODS AND FOUR (4) CLICKX LOCKING CAPS. THE SURGEON WILL MONITOR THE EFFECTIVENESS OF THE CLICK COLUMN SYSTEM THROUGH POSTOPERATIVE RADIOGRAPHIC CONTROLS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CURRENTLY, THE PATIENT WAS IN GOOD CONDITION. THIS COMPLAINT INVOLVES TEN (10) DEVICES.
Patient problems: No Code Available; Pain
MAUDE report 8587489 ↗InjuryJun 11, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED PATIENT UNDERWENT A PROCEDURE ON (B)(6) 2018 TO REMOVE THE 6.2MM PREASSEMBLED CLICKX PEDICLE SCREW DUE TO THE RELEASE OF THE 3D HEAD OF THE SCREW. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) 6.2MM CLICKX PEDICLE SCREW PREASSEMBLED 50MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant; No Code Available
MAUDE report 7589717 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page