SYNFIX Evolution 03.835.100Recalls & reported problems
Synthes GmbH · SYNFIX® EVOLUTION TRIAL SPACER HOLDER · Class 1
All models of SYNFIX Evolution (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034814087), received between Jul 9, 2019 and Apr 1, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility3
- Compatibility Problem2
- Device Difficult to Maintain1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2026Compatibility Problem
IT WAS REPORTED THAT THE SCREWDRIVER, INTENDED TO CONNECT TO THE SYNFIX SPACER TRIALS, WOULD NOT SEAT FLUSH WITH THE TRIALS AND THREADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24758446 ↗MalfunctionJan 27, 2021Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE SYNFIX EVOLUTION TRIAL MEDIUM 15MM WOULD NOT DISENGAGE FROM THE EVOLUTION SPINDLE. THE TRIAL WOULD UNWIND TO DISENGAGE THE KNOB, BUT THE LONG SHAFT WAS STUCK AT THE THREADED PORTION OF THE SPINDLE. THERE WAS NO REPORTED SURGICAL DELAY. THERE WERE NO FRAGMENTS GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORTED CONSEQUENCE TO THE PATIENT. CONCOMITANT DEVICES REPORTED: SYNCAGE EVOLUTION SPINDLE (PART NUMBER 03.825.002, LOT UNKNOWN, QUANTITY 1) THIS REPORT INVOLVES ONE (1) SYNFIX® EVOLUTION TRIAL SPACER HOLDER. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11235932 ↗MalfunctionFeb 5, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(4) AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THE SURGEON WAS USED TRAILED IMPLANT BUT THE SYNFIX EVOLUTION TRIAL MEDIUM HEIGHT WOULD NOT COME OFF THE TRIAL IMPLANT HOLDER. WHEN THE NURSE WAS TRYING TO DISENGAGE THE TRIAL FROM THE EVOLUTION SPINDLE THAT GOES DOWN THE TRIAL IMPLANT HOLDER TO ENGAGE THE TRAIL WOULD UNWIND TO DISENGAGE THE KNOB HOWEVER THE LONG SHAFT WAS STUCK AT THE THREADED PORTION OF THE SPINDLE. THE IMPLANT TRIAL IS STUCK ONTO THE IMPLANT HOLDER (2-PIECE INSTRUMENT) IN SIMPLER TERMS. IT COULD NOT DISENGAGE THE TRIAL FROM THE TRIAL HOLDER. THERE WAS 30 MINUTES DELAY OF SURGERY AND COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE. THIS IS THE REPORT OF (1) SYNFIX EVOL TRIAL IMPLANT HOLDER. THIS IS REPORT 01 OF 03 OF (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9671155 ↗MalfunctionDec 9, 2019Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2019, AT THE END OF THE CASE, I WAS PACKING UP THE SET AND THE SYNFIX EVOLUTION TRIAL M 13.5 10 DEGREE WOULD NOT COME OFF THE TRIAL IMPLANT HOLDER. WHEN I WAS TRYING TO DISENGAGE THE TRIAL THE EVOLUTION SPINDLE THAT GOES DOWN THE TRIAL IMPLANT HOLDER TO ENGAGE THE TRAIL WOULD UNWIND TO DISENGAGE THE KNOB. HOWEVER, THE LONG SHAFT WAS STUCK AT THE THREADER PORTION OF THE SPINDLE. THE IMPLANT TRIAL IS STUCK ONTO THE IMPLANT HOLDER (2 PIECES INSTRUMENT) IN SIMPLER TERMS. NO PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS IS 1 OF 2 FOR REPORT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9437314 ↗MalfunctionSep 12, 2019Device-Device Incompatibility; Physical Resistance/Sticking; Device-Device Incompatibility; Physical Resistance/Sticking
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT AFTER PROCEDURE, THE SURGEON TRIED TO REMOVE TRIAL FROM IMPLANT HOLDER. THE TRIAL WAS STUCK AND COULD NOT BE REMOVED, THE ITEM WAS TAGGED. PATIENT STATUS IS UNKNOWN. THE PRODUCT INVESTIGATION WILL BE COMPLETED BY THE INVESTIGATION TEAM DUE TO THE PROVIDED IMAGES. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9013908 ↗MalfunctionJul 9, 2019Device Difficult to Maintain; Device Difficult to Maintain
07/11/2019: UPDATE: IT WAS REPORTED THAT THE REASON WHY THE PATIENT UNDERWENT THE PROCEDURE WAS TO REINSTATE LORDOSIS INTO THE LUMBAR SPINE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY REPORTED. UPDATED CONCOMITANT DEVICE/S REPORTED TO CAPTURE THE NUMBER OF TRIAL IMPLANTS (3) USED. CONCOMITANT DEVICE/S REPORTED: UNKNOWN TRIAL IMPLANTS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 3).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8772814 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page