The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034817156), received between Dec 24, 2019 and Jan 8, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Migration1
- Patient Device Interaction Problem1
Patient problems
- Joint Laxity3
- Insufficient Information1
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: FAILURE TO REACH THERAPEUTIC GOALS, HARDWARE REMOVAL, IMPLANT FAILURE, SPINAL IMBALANCE. COMPLICATION TIME/INTERVENTION. REPEAT SURGERY. THIS IS FOR DEPUY SYNTHES SPINE T-PAL TI, MATRIX PERFORATED, MATRIX /OTHER AND MATRIX PRODUCTS (T-PAL TI SPACER 10MM X 28MM 10MM HT - ST (B)(4)).
Patient problems: Insufficient Information; Joint Laxity
MAUDE report 18469575 ↗InjuryJan 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: INSTABILITY. COMPLICATION TIME/INTERVENTION. REPEAT SURGERY. THIS IS FOR DEPUY SYNTHES SPINE T-PAL TI PRODUCTS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Joint Laxity
MAUDE report 18468406 ↗InjuryJan 8, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: IMPLANT MALPOSITION, INSTABILITY, NON-UNION, POSTOPERATIVE INFECTION DEEP, COMPLICATION TIME/INTERVENTION, REPEAT SURGERY. THIS REPORT IS FOR A T-PAL TI SPACER 10MM X 28MM 10MM HEIGHT ¿ STERILE. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Joint Laxity; Nonunion/Delayed-union Bone Fracture; Unspecified Infection
MAUDE report 18467636 ↗InjuryDec 24, 2019Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT A REVISION PROCEDURE TO REMOVE THE TPAL (TRANSFORAMINAL POSTERIOR ATRAUMATIC LUMBAR) TITANIUM SPACER DUE TO THE REPORTED T-PAL SPACER THAT HAD MIGRATED OUT OF THE L4-L5 DISC SPACE. INITIALLY, THE PATIENT UNDERWENT AN L3-L5 TLIF POSTERIOR SPINAL FUSION PROCEDURE ON (B)(6) 2019. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THE SURGICAL OUTCOME IS UNKNOWN BUT WITH A PATIENT CONSEQUENCE. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 9517076 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page