Quick summary

This Ethicon Endopath PN120 is a pneumoperitoneum (Veress) insufflation needle used to access the abdominal cavity for minimally invasive surgery. Most reports describe the spring-loaded inner mechanism failing to spring back or sticking when the needle is activated, and a few mention units arriving already opened or non-sterile, with some bleeding and unspecified tissue injury noted. There are 37 reports, including 1 death and 4 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
1
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10705036012962), received between Dec 17, 2019 and Jul 1, 2026 — classified as 1 death, 4 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 8 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available26
  • Insufficient Device Problem Information4
  • Patient Device Interaction Problem4
  • Delivered as Unsterile Product3
  • Improper Flow or Infusion2
  • Mechanical Jam1
  • Retraction Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • Hemorrhage/Blood Loss/Bleeding3
  • No Consequences Or Impact To Patient2
  • Unspecified Tissue Injury2

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT PRIOR TO A ROBOTIC GALLBLADDER PROCEDURE, THE NEEDLE WAS TESTED BEFORE THE CASE DUE TO PREVIOUS ISSUES. DURING TESTING, THE NEEDLE DID NOT SPRING BACK AND APPEARED TO STICK TO THE PLASTIC, BECOMING STUCK. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25683410 ↗
  • MalfunctionMay 23, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A DA VINCI NEPHRECTOMY SURGERY, AFTER ANESTHESIA, WAS PREPARING TO ESTABLISH AN AIRPLANE, AND WHEN USING THE PN120 PNEUMOPERITONEUM NEEDLE, FOUND THAT THE PRODUCT'S INNER SPRING COULD NOT REBOUND BACK CORRECTLY. AFTER REPLACING A NEW ONE, IT IS NORMAL, AND THE PATIENT IS NOT AFFECTED IN ANY WAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25264684 ↗
  • MalfunctionMay 22, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, SPRING MECHANISM ON THE TOP OF THE 120MM VERRIS INSUFFLATION NEEDLE WAS NOT SPRINGING BACK. OPENED A SECOND ONE OF A DIFFERENT LOT NUMBER, SAME ISSUE. SWITCHED TO 150MM VERRIS TO CONTINUE CASE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256039 ↗
  • MalfunctionMar 4, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, HAD ISSUES WITH THE SPRING IN IT. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24506815 ↗
  • MalfunctionOct 21, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE SPRING OF THE VERESS NEEDLE FAILED TO RETRACT. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23346523 ↗
  • MalfunctionSep 8, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN CHECKING THE BOX, AN OPEN PIECE IS FOUND WITHOUT BEING ABLE TO BE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22990860 ↗
  • MalfunctionJul 7, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A HERNIA REPAIR THE NURSE INSTRUMENTALIST CHECKS THE NEEDLE BEFORE STARTING THE PROCEDURE AND REALIZES THAT WHEN IT IS ACTIVATED IT DOES NOT RETURN. THE DOCTOR REVIEWS IT AND GIVES THE SAME VERDICT. BY NOT RETURNING THE NEEDLE TO THE ORIGINAL POSITION ALONE, IT WOULD NOT WORK TO DETERMINE IF THE DOCTOR ENTERS THE CAVITY CORRECTLY. ANOTHER IS THEREFORE REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22411260 ↗
  • MalfunctionApr 23, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE INSUFFLATION NEEDLE FAILED TO RETRACT. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21893185 ↗
  • MalfunctionApr 2, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING IS NOT FUNCTIONALLY PROPERLY. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751021 ↗
  • MalfunctionApr 2, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING IS NOT FUNCTIONALLY PROPERLY. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751020 ↗
  • MalfunctionApr 2, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING IS NOT FUNCTIONALLY PROPERLY. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751018 ↗
  • MalfunctionMar 25, 2025Improper Flow or Infusion; Appropriate Device Problem Term/Code Not Available; Improper Flow or Infusion; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, NEEDLE IS BLOCKED AND THE SAFETY SYSTEM IS FAULTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687573 ↗
  • MalfunctionNov 25, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE NEEDLE WAS HAVING TROUBLE SPRINGING BACK UP CORRECTLY AND WAS GETTING STUCK. HAD TO GET A NEW ONE. NO PATIENT CONSEQUENCE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20775252 ↗
  • MalfunctionOct 24, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING CANNOT REBOUND. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20525212 ↗
  • MalfunctionOct 23, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING CANNOT BOUNCE BACK. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20515807 ↗
  • MalfunctionOct 22, 2024Improper Flow or Infusion; Appropriate Device Problem Term/Code Not Available; Improper Flow or Infusion; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE PNEUMONEEDLE WAS INSERTED INTO THE INCISION SITE, BUT THE SPRING DID NOT WORK AS INTENDED TO GIVE PHYSICAL FEEDBACK OF ABDOMINAL LAYERS. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20508005 ↗
  • MalfunctionOct 2, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SAFETY NEEDLE REMAINED EXPOSED AFTER PUNCTURE AND DID NOT RETRACT. OPENED A NEW PRODUCT TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20351295 ↗
  • MalfunctionOct 2, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SAFETY NEEDLE REMAINED EXPOSED AFTER PUNCTURE AND DID NOT RETRACT. OPENED A NEW PRODUCT TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20351291 ↗
  • MalfunctionOct 2, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN ROBOTIC PROCEDURE, THE SAFETY NEEDLE REMAINED EXPOSED AFTER PUNCTURE AND DID NOT RETRACT. OPENED A SECOND PRODUCT TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20351135 ↗
  • MalfunctionOct 2, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A ROBOTIC NEPHRECTOMY, THE SAFETY NEEDLE REMAINED EXPOSED AFTER PUNCTURE AND DID NOT RETRACT. OPENED A NEW PRODUCT TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20350998 ↗
  • MalfunctionOct 2, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A ROBOTIC NEPHRECTOMY, THE SAFETY NEEDLE REMAINED EXPOSED AFTER PUNCTURE AND DID NOT RETRACT. OPENED A NEW PRODUCT TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20350993 ↗
  • MalfunctionAug 22, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING IS STUCK AND CANNOT REBOUND. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20034324 ↗
  • MalfunctionJul 16, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING CANNOT BOUNCE BACK. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19757038 ↗
  • MalfunctionJun 11, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT PRE PROCEDURE, IT WAS NOTICED THAT THERE WAS A HOLE IN THE BAG, RENDERING IT NONSTERILE. CASE COMPLETED WITH LIKE DEVICE. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19512849 ↗
  • MalfunctionMar 19, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SPRING COULD NOT REBOUND. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18932062 ↗

FDA clearance and classification

Cleared through 510(k) K910875 on Mar 19, 1991 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Endopath Disposable Pneumoperitoneum Needle”. The FDA review took 18 days. FDA record ↗

The FDA regulates product code FDP (apparatus, pneumoperitoneum, automatic) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Endopath PN120

Has the Endopath PN120 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036012962), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Endopath PN120?

The FDA's MAUDE database lists 36 adverse event reports for this device, received between Dec 17, 2019 and Jul 1, 2026. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (4), “Delivered as Unsterile Product” (3) and “Improper Flow or Infusion” (2); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (3) and “Unspecified Tissue Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Endopath PN120?

1 report is classified as a death, 4 as injuries and 31 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Endopath PN120 FDA approved?

It was cleared through 510(k) K910875 on Mar 19, 1991. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Endopath PN120, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand Endopath, model PN120). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other APPARATUS, PNEUMOPERITONEUM, AUTOMATIC devices

All APPARATUS, PNEUMOPERITONEUM, AUTOMATIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page