Skip to content

STERRAD™ 10104-005Recalls & reported problems

Advanced Sterilization Products Services Inc. · STERRAD™ 100NX Sterilization System with ALLClear™ Technology, 1-Door · Class 2

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705037048670), received between Feb 28, 2019 and Apr 10, 2023 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Environmental Particulates” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Environmental Particulates7
  • Device Emits Odor4
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient4
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2023Device Emits Odor; Device Emits Odor

    A CUSTOMER REPORTED AN EVENT OF AN ¿ODOR¿ OR SMELL EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A PREVIOUS SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16710341 ↗
  • MalfunctionJul 8, 2022Device Emits Odor; Device Emits Odor

    A CUSTOMER REPORTED AN EVENT OF A ¿SMELL¿ OR ODOR EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14956803 ↗
  • MalfunctionApr 18, 2022Environmental Particulates; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF AN ¿OIL MIST¿ OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14138167 ↗
  • MalfunctionApr 8, 2022Environmental Particulates; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF A ¿SMOKE¿ OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14061265 ↗
  • MalfunctionFeb 3, 2022Environmental Particulates; Environmental Particulates

    A BIOMED REPORTED AN EVENT OF A ¿SMOKE¿ OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP SERVICE ENGINEER PROVIDED REMOTE SUPPORT AND INSTRUCTED THE BIOMED TO ADJUST THE OIL MIST FILTER TO RESOLVE THE SMOKE/HAZE ISSUE. THE STATUS OF THE UNIT WAS REPORTED AS ¿NOW SYSTEM IS FUNCTIONING NORMALLY.¿ THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13446912 ↗
  • MalfunctionJun 2, 2021Environmental Particulates; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF A "SMOKE" OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: Insufficient Information

    MAUDE report 11925908 ↗
  • MalfunctionOct 28, 2020Device Emits Odor; Smoking

    A CUSTOMER REPORTED AN EVENT OF A "FUMES" OR ODOR/SMELL/SMOKE/HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10747813 ↗
  • MalfunctionMar 17, 2020Environmental Particulates; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF A "SMOKE" OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9845491 ↗
  • MalfunctionNov 26, 2019Device Emits Odor; Environmental Particulates; Device Emits Odor; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF "FUMES" EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9376302 ↗
  • MalfunctionFeb 28, 2019Environmental Particulates; Environmental Particulates

    A CUSTOMER REPORTED AN EVENT OF A "WHITE SMOKE" OR HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8380021 ↗

Other Sterilizer, chemical devices

All Sterilizer, chemical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.