Skip to content

Hemochron Whole Blood Microcoagulation System JACT+Recalls & reported problems

Accriva Diagnostics Holdings, Inc. · ACT+ Cuvette - 45 cuvettes/box. · Class 2

All models of Hemochron Whole Blood Microcoagulation System (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10711234103067), received between Sep 24, 2024 and Sep 24, 2024 — classified as 1 death.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Ischemic Heart Disease1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathSep 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED THAT A HEMOCHRON SIGNATURE ELITE ANALYZER AND ACT+ TEST DEVICE WAS USED TO MONITOR HEPARIN ANTICOAGULATION IN A PATIENT THAT WAS UNDERGOING A CORONARY ANGIOPLASTY PROCEDURE. ACT TEST RESULT OBTAINED AT BASELINE WAS NOT SPECIFIED. THE PATIENT RECEIVED CLOPIDOGREL BY MOUTH AND WAS GIVEN A 5000U INTRAVENOUS DOSE OF HEPARIN. THE RESULT OF AN ACT TEST OBTAINED 1 HOUR AFTER HEPARIN ADMINISTRATION WAS OUT-OF-RANGE HIGH. THE PATIENT EXPIRED DUE TO A SLOW HEART RATE AT AN UNSPECIFIED TIME. NO OTHER MEDICAL OR LABORATORY INFORMATION WAS SPECIFIED. ELECTRONIC QUALITY CONTROL (EQC) TESTING PERFORMED PRIOR TO THE PROCEDURE PASSED.

    Patient problems: Ischemic Heart Disease

    MAUDE report 20295483 ↗

Other Activated Whole Blood Clotting Time devices

All Activated Whole Blood Clotting Time devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page