Skip to content

Avoximeter Whole Blood CO-Oximeter AVOX4000Recalls & reported problems

Accriva Diagnostics Holdings, Inc. · Avoximeter Whole Blood CO-Oximeter System - Instrument · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10711234530054), received between Jan 8, 2018 and Jan 8, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K951485); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2018High Test Results; High Test Results

    HEALTHCARE PROFESSIONAL REPORTED AN UNEXPECTEDLY HIGH %COHB RESULT WITH AN AVOXIMETER 4000 CO-OXIMETER IN A CONTINUOUSLY VENTILATED NURSING HOME PATIENT. BLOOD GASES AND CO-OXIMETRY TESTS WERE BEING CONDUCTED EVERY 12 HOURS. TEST RESULTS WERE AS EXPECTED EXCEPT FOR ONE UNEXPECTEDLY HIGH %COHB RESULT (5.2%) THAT WAS CLINICALLY SIGNIFICANT AND ABOVE THE PATIENT'S 48-HOUR BASELINE %COHB OF 2.1%. THE PATIENT RECEIVED AN UNSPECIFIED TREATMENT AND SUBSEQUENT BLOOD GASES AND CO-OXIMETER RESULTS RETURNED TO BASELINE. AN INTERNAL INVESTIGATION DID NOT FIND ANY EVIDENCE THAT THE PATIENT WAS EXPOSED TO CARBON MONOXIDE. NO MEDICAL COMPLICATIONS OR OTHER EVENTS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7173967 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K951485. One submission often covers many models, so check the product and lot details in each record.

  • Z-2218-2011 · Mar 23, 2011 · Terminated
    AVOXimeter 1000, AVOXimeter 4000 1000- portable bedside whole blood oximeter that directly measures oxyhemoglobin fraction, the total hemoglobin concentration, and oxygen content.
  • Z-1239-2008 · Feb 13, 2008 · Terminated
    International Technidyne Corp. (ITC) AVOXimeter 4000 Co-oximeter, Catalog Numbers: AVOX4000-110 and AVOX4000-220

Other Spectral Absorb. Curve, Oxyhemoglobin, Carboxyhemoglobin, Carbon-Monoxide devices

All Spectral Absorb. Curve, Oxyhemoglobin, Carboxyhemoglobin, Carbon-Monoxide devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page