Bausch & Lomb Incorporated · SofPort Silicone Advanced Optics Aspheric Lens +15.50D · Class 3 · Implant
All models of SofPort IOL (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770474223), received between Nov 15, 2023 and Nov 15, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 15, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO EYE, THE SURGEON NOTICED THE FRONT HAPTIC WAS KINKED. THE INCISION WAS ENLARGED TO ALLOW FOR REMOVAL OF THE IOL AND THE LENS WAS EXCHANGED WITH A DIFFERENT MODEL IOL. SUTURES WERE USED TO CLOSE THE WOUND. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF EVENT WAS A BAD LOAD, REPORTING THE TECH ADVANCED THE TAB TOO FAST. ADDITIONAL INFORMATION WAS REQUESTED, BUT NOT RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18145394 ↗
Other Intraocular Lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page