Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10757770474285), received between Feb 15, 2023 and Apr 22, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Material Split, Cut or Torn1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Capsular Bag Tear1
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING INSERTION OF AN INTRAOCULAR LENS (IOL), THE HAPTIC BECAME STUCK IN THE INJECTOR AND WAS AMPUTATED. THE LENS WAS REMOVED AND THE INCISION WAS ENLARGED. A REPLACEMENT IOL OF THE SAME MODEL AND POWER WAS PLACED, AND A SUTURE WAS REQUIRED. THE PATIENT IS EXPECTED TO EXPERIENCE A PROLONGED RECOVERY PERIOD; HOWEVER, THE OVERALL PROGNOSIS IS EXCELLENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24955880 ↗
  • InjuryNov 18, 2024Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THREE DAYS AFTER THE IMPLANTATION OF AN INTRALOCULAR LENS(IOL), INTO THE RIGHT EYE (OD), A SUCCESSFUL LENS EXCHANGE USING A DIFFERENT MODEL IOL WAS COMPLETED DUE TO DISLOCATION OF THE IOL. A VITRECTOMY WAS PERFORMED. ADDITIONAL INFORMATION REQUESTED, BUT NOT RECEIVED.

    Patient problems: Insufficient Information

    MAUDE report 20717523 ↗
  • InjurySep 7, 2023Unintended Movement; Unintended Movement

    IT WAS REPORTED DURING IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE, THE POSTERIOR CAPSULE RUPTURED CAUSING THE LENS TO BE UNSTABLE. THE INCISION WAS ENLARGED TO ALLOW FOR REMOVAL OF THE IOL AND A SUTURE WAS PLACED AS STANDARD PROTOCOL. APPROXIMATELY TEN DAYS AFTER EXPLANT, A VITRECTOMY WAS PERFORMED AND A LENS OF THE SAME MODEL, SAME DIOPTER WAS IMPLANTED. PATIENT OUTCOME IS GOOD.

    Patient problems: Capsular Bag Tear

    MAUDE report 17702006 ↗
  • InjuryApr 12, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE LENS WAS REMOVED AND REPLACED INTRAOPERATIVELY DUE TO A BENT HAPTIC. THE INCISION WAS ENLARGED. NO SUTURES WERE REQUIRED AND THERE WAS NO IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NOT RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16724560 ↗
  • InjuryFeb 15, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE EYE, IT WAS NOTICED THE HAPTIC WAS DAMAGED, ALLEGEDLY BY THE SLIDE MECHANISM OF THE INJECTOR. THE INCISION WAS ENLARGED AND A SUCCESSFUL INTRAOPERATIVE LENS REPLACEMENT USING THE SAME MODEL AND DIOPTER IOL WAS COMPLETED. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT NOT RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16380622 ↗