Skip to content

Trulign® Toric IOL BL1UTRecalls & reported problems

Bausch & Lomb Incorporated · Trulign toric posterior chamber intraocular lens +25.50D SE / +2.75D CYL · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770521392), received between Sep 8, 2023 and Sep 8, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Blurred Vision1
  • Capsular Contracture1
  • UGH (Uveitis Glaucoma Hyphema) Syndrome1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT APPROXIMATELY TEN WEEKS AFTER IMPLANTATION AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE, POSTERIOR AND ASYMMETRIC LENS VAULTING WAS OBSERVED. IT ALSO WAS REPORTED THE PATIENT DEVELOPED UVEITIS GLAUCOMA HYPHEMA (UGH) SYNDROME. THE IOL WAS EXCHANGED WITH A DIFFERENT MODEL DUE TO UNRESOLVED Z-SYNDROME AND CAPSULAR PHIMOSIS WITHIN A WEEK OF ONSET OF SYMPTOMS. YAG CAPSULOTOMY WAS PERFORMED ONE DAY POST IOL EXCHANGE. IN THE EXPLANTING SURGEON''S OPINION, THE LIKELY CAUSE OF THE EVENT IS CAPSULAR CONTRACTION. IN THE IMPLANTING SURGEON''S OPINION, THE MOST LIKELY CAUSE OF THE EVENT IS SIZE OF THE LENS RELATIVE TO THE SULCUS DIAMETER, CAPSULAR FIBROSIS, AND CAPSULAR PHIMOSIS. PATIENT OUTCOME IS GOOD.

    Patient problems: Blurred Vision; Capsular Contracture; UGH (Uveitis Glaucoma Hyphema) Syndrome

    MAUDE report 17712135 ↗

Other Lens, Intraocular, Toric Optics devices

All Lens, Intraocular, Toric Optics devices →