enVista® IOL MX60E (REF MXUE0100)Recalls & reported problems
Bausch & Lomb Incorporated · enVista one-piece hydrophobic acrylic intraocular lens +1.00D · Class 3 · Implant
All models of enVista IOL (74)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770530967), received between Apr 30, 2025 and Apr 30, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- Toxic Anterior Segment Syndrome (TASS)1
- Visual Impairment1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT AFTER ONE DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE EYE, THE PATIENT PRESENTED WITH INTENSE INFLAMMATION AND THICK FIBROTIC TISSUE AROUND THE LENS. THE PATIENT WAS REFERRED TO A RETINA SPECIALIST AND TREATED WITH FREQUENT STEROIDS. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). IN THE SURGEON¿S OPINION, THE MOST LIKELY CAUSE OF THE EVENT IS AN ATYPICAL PRESENTATION OF TASS. ADDITIONAL INFORMATION WAS REQUESTED, BUT NOT RECEIVED.
Patient problems: Toxic Anterior Segment Syndrome (TASS); Visual Impairment
MAUDE report 21933846 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page