enVista® IOL EE (REF EEU2100)Recalls & reported problems
Bausch & Lomb Incorporated · enVista hydrophobic acrylic IOL (non-preloaded) +21.00D · Class 3 · Implant
All models of enVista IOL (74)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10757770607898), received between Apr 18, 2025 and Apr 23, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Class IOpen, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1679-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information3
Patient problems
- Toxic Anterior Segment Syndrome (TASS)3
- Visual Impairment2
- Corneal Edema1
- Endophthalmitis1
- Hypopyon1
- Keratitis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT 1 DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE, THE PATIENT PRESENTED WITH DECREASED VISUAL ACUITY, INCREASED IOP (PRE-OP 14 MMHG AND POST OP 22 MMHG), HYPOPYON AND POSSIBLE ENDOPHTHALMITIS. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). CULTURES WERE TAKEN AND THE RESULTS WERE NEGATIVE. THE PATIENT WAS TREATED WITH INTRAOCULAR INJECTIONS OF CEFTAZIDINE AND VANCOMYCIN, SUB TENONS KENALOG, DUREZOL AND OFLOXACIN. IN THE SURGEON¿S OPINION, THE MOST LIKELY CAUSE OF THE EVENT WAS RELATED TO THE IMPLANTED IOL. PATIENT OUTCOME WAS REPORTED TO BE GOOD.
Patient problems: Endophthalmitis; Hypopyon; Toxic Anterior Segment Syndrome (TASS); Visual Impairment
MAUDE report 21896675 ↗InjuryApr 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT ONE DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE, THE PATIENT PRESENTED WITH POOR VISION WITH NO PAIN. SLIT LAMP FINDINGS WERE 1+ EDEMA, 4+ CELL WITH FIBRIN, 1+ SPK NUCLEAR FRAGMENT. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE FRAGMENT WAS REMOVED AND THE PATIENT'S OUTCOME IS GOOD. IN THE SURGEON¿S OPINION, THE MOST LIKELY CAUSE OF THE EVENT TASS. THE FOLLOWING MEDICATIONS WERE PRESCRIBED (FOR USE AT HOME POST-OPERATIVELY): PRED/MOXI/NEPA, PRED ACETATE AND MEDROL DOSE PACK.
Patient problems: Corneal Edema; Keratitis; Toxic Anterior Segment Syndrome (TASS); Visual Impairment
MAUDE report 21896030 ↗InjuryApr 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ADDITIONAL INFORMATION RECEIVED. THE PATIENT'S BEST CORRECTED VISION FOR FINAL OUTCOME WAS 20/20 IN THE AFFECTED EYE.
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 21865609 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page