Skip to content

enVista Aspire™ IOL EA (REF EAU2150)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Aspire hydrophobic acrylic IOL (non-preloaded) +21.50D · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10757770608581), received between Apr 24, 2025 and Oct 22, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Open, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1679-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Capsular Bag Tear1
  • Toxic Anterior Segment Syndrome (TASS)1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING INSERTION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE (OD) THE CAPSULE RUPTURED, AND THE IOL DROPPED THROUGH THE DEFECT IN THE CAPSULAR BAG. THE IOL WAS REMOVED, A VITRECTOMY WAS PERFORMED, AND A DIFFERENT MODEL IOL WAS IMPLANTED INTO THE SULCUS. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT WAS THE IOL DROPPED THROUGH CAPSULAR DEFECT. THE PATIENT'S OUTCOME IS GOOD.

    Patient problems: Capsular Bag Tear

    MAUDE report 23360956 ↗
  • InjuryApr 24, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT ONE DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE, THE PATIENT PRESENTED WITH CELL AND FIBRIN. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE PATIENT WAS TREATED WITH MOXI, BROM AND PRED AND AN ANTIBIOTIC INJECTION. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT WAS TASS. PATIENT OUTCOME WAS REPORTED TO BE GOOD.

    Patient problems: Toxic Anterior Segment Syndrome (TASS)

    MAUDE report 21902622 ↗

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page