enVista® Toric IOL ETE (REF ETEU300+240)Recalls & reported problems
Bausch & Lomb Incorporated · enVista toric hydrophobic acrylic IOL (non-preloaded) +24.00D SE / +3.00D CYL · Class 3 · Implant
All models of enVista Toric IOL (450)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770614063), received between Sep 29, 2026 and Sep 29, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Class IOpen, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1679-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- Hypopyon1
- Toxic Anterior Segment Syndrome (TASS)1
- Visual Impairment1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Insufficient Device Problem Information
It was reported that approximately five weeks after implantation of an intraocular lens (IOL), the patient presented with fibrin and hypopyon in the anterior chamber. Visual acuity decreased to count fingers at 20 feet. Based on the clinical findings the surgeon concluded the cause as Toxic Anterior Segment Syndrome (TASS). A Tap procedure was performed, and a corticosteroid injection was administered. Due to the severity of the inflammation, oral corticosteroid therapy was initiated. The patient's condition subsequently improved; however, vision remains significantly reduced. Additional information was requested but was not received.
Patient problems: Hypopyon; Toxic Anterior Segment Syndrome (TASS); Visual Impairment
MAUDE report 50931606 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page