Skip to content

enVista Aspire™ Toric IOL ETA (REF ETAU350+085)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Aspire toric hydrophobic acrylic IOL (non-preloaded) +8.50D SE / +3.50D CYL · Class 3 · Implant

All models of enVista Aspire Toric IOL (513)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 27, 2025Class IOpen, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1679-2025 · Root cause: Process control · Recall details · FDA record ↗

Other Intraocular Lens devices

All Intraocular Lens devices →