Skip to content

enVista Envy™ IOL (REF ENU1000)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Envy hydrophobic acrylic IOL (non-preloaded) +10.00D · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770637079), received between Apr 7, 2026 and Apr 7, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Open, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Corneal Edema1
  • Hypopyon1
  • Posterior Capsule Opacification1
  • Toxic Anterior Segment Syndrome (TASS)1
  • Visual Impairment1
  • Vitreous Floaters1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 7, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT 1 DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE (OD), THE PATIENT PRESENTED AT THE POSTOPERATIVE APPOINTMENT WITH THE CHIEF COMPLAINT OF AN INABILITY TO SEE. SLIT LAMP EXAMINATION IDENTIFIED 4+ CORNEAL EDEMA. INTRAVITREAL INJECTIONS OF VANCOMYCIN AND CEFTAZIDIME WERE ADMINISTERED THE SAME DAY. CULTURE AND SENSITIVITY TESTING REVEALED NO ORGANISMS, WITH MODERATE WHITE BLOOD CELLS PRESENT. THE RETINA SPECIALIST, WHO MANAGED THE PATIENT POSTOPERATIVELY, DOCUMENTED A DIFFERENTIAL DIAGNOSIS OF TOXIC ANTERIOR SEGMENT SYNDROME (TASS) VERSUS ENDOPHTHALMITIS. AT THE 1-WEEK POSTOPERATIVE VISIT, SLIT LAMP FINDINGS SHOWED IMPROVEMENT, FIBRIN MEMBRANE AND HYPOPYON WERE NO LONGER PRESENT. AT THE 2-WEEK POSTOPERATIVE VISIT, THE PATIENT REPORTED INTERMITTENT BLURRY VISION. CLINICAL EXAMINATION CONTINUED TO SHOW IMPROVEMENT, INCLUDING DECREASED VITREOUS DEBRIS AND A CLEAR CORNEA. AT APPROXIMATELY SEVEN WEEKS POSTOPERATIVE, THE PATIENT DEMONSTRATED CONTINUED CLINICAL IMPROVEMENT, THE VITREOUS WAS NOTED TO BE CLEAR WITHOUT HEMORRHAGE, CELLS, OR PIGMENT. THE PATIENT WAS DIAGNOSED WITH POSTERIOR CAPSULE OPACIFICATION (PCO). AT 4.5 MONTHS POSTOPERATIVE, THE

    Patient problems: Corneal Edema; Hypopyon; Posterior Capsule Opacification; Toxic Anterior Segment Syndrome (TASS); Visual Impairment; Vitreous Floaters

    MAUDE report 24804434 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page