enVista Envy™ IOL (REF ENU1150)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Envy hydrophobic acrylic IOL (non-preloaded) +11.50D · Class 3 · Implant

All models of enVista Envy IOL (53)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770637109), received between Jun 18, 2025 and Jun 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Open, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • Hypopyon1
  • Intraocular Pressure Increased1
  • Toxic Anterior Segment Syndrome (TASS)1
  • Visual Impairment1
  • Vitreous Floaters1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT 1 DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE (OS), THE PATIENT PRESENTED WITH 3+ CELLS, FIBRIN, AND HYPOPYON. PATIENT'S BEST CORRECTED DISTANCE VISUAL ACUITY (BCDVA) DECREASED FROM 20/20-1 TO 20/60-1. BASED ON THE FINDINGS, THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS) AND INCREASED STEROID DROPS. AT THE PATIENT'S 1-MONTH POST-OP VISIT, INFLAMMATION HAD RESOLVED WITH UNCORRECTED DISTANCE VISUAL ACUITY (UCDVA) OF 20/20 AND UNCORRECTED NEAR VISUAL ACUITY (UCNVA) OF J1. DUE TO FINDINGS OF VITREOUS DEBRIS/VEIL, THE PATIENT WAS REFERRED TO RETINA FOR AN EVALUATION. APPROXIMATELY ONE WEEK LATER, A PARS PLANA VITRECTOMY (PPV) WAS PERFORMED DUE TO RESIDUAL VITREOUS DEBRIS. ON WEEK AFTER PPV, THE VITREOUS WAS FOUND TO BE CLEAR. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT IS THE IOL. PATIENT OUTCOME IS GOOD. SURGEON WILL CONTINUE TO MONITOR INTRAOCULAR PRESSURE. MEDICATIONS PRESCRIBED FOR USE AT HOME POST-OPERATIVELY: PRED ACETATE 1% 1 GTT HOURLY. COSOPT 1 GTT BID (ADDED 21-MAR-2025). BRIMONIDINE 0.2% 1GTT BID (ADDED 22-MAR-2025).

    Patient problems: Hypopyon; Intraocular Pressure Increased; Toxic Anterior Segment Syndrome (TASS); Visual Impairment; Vitreous Floaters

    MAUDE report 22287524 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →