enVista Envy™ IOL (REF ENU1550)Recalls & reported problems
Bausch & Lomb Incorporated · enVista Envy hydrophobic acrylic IOL (non-preloaded) +15.50D · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770637178), received between Aug 21, 2026 and Aug 21, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Wrong Label” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Open, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1680-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Wrong Label1
Patient problems
- Visual Impairment1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 21, 2026Wrong Label
It was reported that after implantation of an intraocular lens (IOL) in the right eye (OD), the refractive outcome revealed an approximately 3.0 diopter refractive miss. The IOL power calculations were reviewed and confirmed to be accurate, indicating that the patient should have achieved the intended refractive outcome. Based on these findings, the surgeon suspected that the implanted IOL may have been mislabeled and elected to perform an IOL exchange using the same model and diopter power approximately 6 weeks after the initial procedure. Patient outcome is good. In the surgeon's opinion, the most likely cause of the event is the initial IOL was mislabeled.
Patient problems: Visual Impairment
MAUDE report 50376656 ↗
Other Lens, Multifocal Intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.