enVista Envy™ IOL (REF ENU1800)Recalls & reported problems
Bausch & Lomb Incorporated · enVista Envy hydrophobic acrylic IOL (non-preloaded) +18.00D · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10757770637222), received between Apr 23, 2026 and Apr 23, 2026 — classified as 2 injuries.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Open, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1680-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Halo2
- Posterior Capsule Opacification2
- Visual Disturbances2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT REFRACTIVE LENS EXCHANGE (RLE) WITH IMPLANTATION OF AN INTRAOCULAR LENS (IOL) IN THE LEFT EYE (OS) AND, APPROXIMATELY TWO MONTHS LATER, COMPLAINED OF HALOS AROUND LIGHTS AND FOGGY VISION. CLINICAL EVALUATION OBSERVED POSTERIOR CAPSULAR OPACIFICATION (PCO). APPROXIMATELY ONE WEEK AFTER SYMPTOM ONSET, THE IOL WAS EXCHANGED, AND THE PATIENT SUBSEQUENTLY RECOVERED. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT WAS MALADAPTATION DUE TO GLARE. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT RECEIVED. THIS IS THE REPORT FOR THE LEFT EYE (OS). RIGHT EYE (OD) REFERENCE: (B)(4).
Patient problems: Halo; Posterior Capsule Opacification; Visual Disturbances
MAUDE report 24965045 ↗InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT REFRACTIVE LENS EXCHANGE (RLE) WITH IMPLANTATION OF AN INTRAOCULAR LENS (IOL) IN THE RIGHT EYE (OD) AND, APPROXIMATELY TWO MONTHS LATER, COMPLAINED OF HALOS AROUND LIGHTS AND FOGGY VISION. CLINICAL EVALUATION OBSERVED POSTERIOR CAPSULAR OPACIFICATION (PCO). APPROXIMATELY ONE WEEK AFTER SYMPTOM ONSET, THE IOL WAS EXCHANGED, AND THE PATIENT SUBSEQUENTLY RECOVERED. IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT WAS MALADAPTATION DUE TO GLARE. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT RECEIVED. THIS IS THE REPORT FOR THE RIGHT EYE (OD). LEFT EYE (OS) REFERENCE: 0001313525-2026-00037
Patient problems: Halo; Posterior Capsule Opacification; Visual Disturbances
MAUDE report 24965011 ↗
Other Lens, Multifocal Intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.