Skip to content

enVista Envy™ IOL (REF ENU2250)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Envy hydrophobic acrylic IOL (non-preloaded) +22.50D · Class 3 · Implant

All models of enVista Envy IOL (53)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10757770637307), received between Apr 18, 2025 and Sep 30, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Open, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Insufficient Device Problem Information1

Patient problems

  • Visual Disturbances2
  • Halo1
  • Insufficient Information1
  • Toxic Anterior Segment Syndrome (TASS)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately four months post intraocular lens (IOL) implantation in the right eye, the lens was explanted due to the patient experiencing glares and halos. A back up lens of a different model was placed.

    Patient problems: Halo; Visual Disturbances

    MAUDE report 50944761 ↗
  • InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately 9 months after implantation of an intraocular lens (IOL) in the right eye (OD), the lens was explanted.This report is for the right eye (OD). Left eye (OS) reference: 0001313525-2026-00124.

    Patient problems: Insufficient Information

    MAUDE report 50429030 ↗
  • InjuryJul 21, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED EIGHT DAYS AFTER A MULTIFOCAL LENS (IOL) WAS PLACED IN THE PATIENTS LEFT EYE (OS) THE PATIENT FELT THEIR QUALITY OF VISION AND COLOR VISION WAS POOR. APPROXIMATELY NINE WEEKS AFTER THE LENS WAS INITIALLY PLACED IT WAS EXPLANTED AND REPLACED FOR A DIFFERENT MODEL LENS. THE SURGEON CITES PATIENT DISSATISFACTION AS THE ISSUE WITH A GOOD POST EXCHANGE OUTCOME.

    Patient problems: Visual Disturbances

    MAUDE report 25898719 ↗
  • InjuryApr 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT 1 DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE (OS), THE PATIENT PRESENTED WITH INCREASED ANTERIOR CHAMBER CELLS & FIBRIN. PATIENT'S BCVA DID NOT DECREASE. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). IN THE SURGEON'S OPINION, THE MOST LIKELY CAUSE OF THE EVENT IS RELATED TO IOL STORAGE MEDIA. PATIENT OUTCOME IS GOOD. THE FOLLOWING MEDICATIONS WERE PRESCRIBED (FOR USE AT HOME POST-OPERATIVELY): TOBRADEX 1 GTT EVERY HOUR, ZYMAMID 1 GTT EVERY HOUR, AVELOX 400G QD 1 WEEK.

    Patient problems: Toxic Anterior Segment Syndrome (TASS)

    MAUDE report 21864665 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →

Sources and updates