enVista Envy™ Toric IOL ETN (REF ETNU125+150)Recalls & reported problems
Bausch & Lomb Incorporated · enVista Envy toric hydrophobic acrylic IOL (non-preloaded) +15.00D SE / +1.25D CYL · Class 3 · Implant
All models of enVista Envy Toric IOL (513)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770640178), received between Jul 14, 2025 and Jul 14, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Class IOpen, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Corneal Clouding/Hazing1
- Corneal Edema1
- Intraocular Pressure Increased1
- Loss of Vision1
- Retinal Detachment1
- Toxic Anterior Segment Syndrome (TASS)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 14, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE PATHOLOGY RESULTS OF THE PRIOR FIBRIN REMOVAL PERFORMED BY THE IMPLANTING SURGEON WAS READ AS FIBRINOUS MEMBRANE WITH NO BACTERIAL AGGREGATES. THE PATIENT WAS REFERRED TO A RETINA SURGEON AGAIN WHERE THE IOL, INCLUDING THE CAPSULAR BAG WAS REMOVED, AS THE MEMBRANE ENCASED THE IOL. AN ANTERIOR VITRECTOMY WAS PERFORMED, CULTURES WERE REPEATED, AND THE PATIENT RECEIVED ANOTHER ROUND OF INTRAVITREAL ANTIBIOTICS. THERE WAS NO GROWTH FROM THE CULTURES. POST-EXPLANT, VISUAL ACUITY WAS LIGHT PERCEPTION (LP) AND INTRAOCULAR PRESSURE WAS VERY LOW. THE PATIENT HAD AN ADDITIONAL SURGERY WITH A PROSTHETIC CORNEA SO THAT THE RETINA COULD BE VISUALIZED AND REPAIR WHAT THEY ASSESSED AS A RETINAL DETACHMENT WITH MACULA OFF. ONCE INTRA-OP, THE SURGEON PERFORMING THIS PROCEDURE FELT THE RETINA WAS NECROTIC AND THERE WAS NO SAVING THE EYE AND PROCEDURE WAS ABORTED. THEIR ASSESSMENT WAS THAT THIS WAS AN AGGRESSIVE STREP INFECTION NOT A TASS. THE MOST RECENT VISUAL ACUITY IS NO LIGHT PERCEPTION (NLP) AND THE EYE IS PHTHISIC. IN THE IMPLANTING SURGEON AND THE RETINA SURGEON¿S (WHO SAW THE PATIENT DAY 2 POST-OP) OPINION, THE MOST LIKELY CAUSE OF THE EVENT
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Corneal Clouding/Hazing; Corneal Edema; Intraocular Pressure Increased; Loss of Vision; Retinal Detachment; Toxic Anterior Segment Syndrome (TASS)
MAUDE report 22499828 ↗
Other Lens, Multifocal Intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page