1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10757770640840), received between Apr 23, 2025 and Apr 21, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Class IOpen, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • Eye Pain1
  • Inflammation1
  • Toxic Anterior Segment Syndrome (TASS)1
  • Visual Disturbances1
  • Visual Impairment1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DYSPHOTOPSIA IN THE RIGHT EYE (OD) FOLLOWING AN INTRAOCULAR LENS (IOL) EXCHANGE. THE PATIENT HAD A FAVORABLE OUTCOME WITH THE MONOFOCAL IOL IN THE FELLOW EYE AND REQUESTED THAT THE RIGHT EYE IOL BE EXCHANGED FOR A MONOFOCAL AS WELL. A SECOND LENS EXCHANGE WAS SUBSEQUENTLY PERFORMED APPROXIMATELY TWO MONTHS AFTER THE FIRST EXCHANGE. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT RECEIVED.

    Patient problems: Visual Disturbances

    MAUDE report 24938271 ↗
  • InjuryApr 23, 2025Manufacturing, Packaging or Shipping Problem; Insufficient Device Problem Information; Manufacturing, Packaging or Shipping Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT APPROXIMATELY 3 WEEKS AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE, THE PATIENT PRESENTED WITH INFLAMMATION. SLIT LAMP FINDINGS WERE 1+ CELL IN A/C AND TRACE FLARE. PATIENT'S BCVA DROPPED FROM 20/25 TO 20/30. PATIENT HAS HAD PERSISTENT INFLAMMATION, EYE PAIN AND HAZY VISION SINCE. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE PATIENT WAS TREATED WITH IMPRISM. FIVE MONTHS AND ONE WEEK AFTER IMPLANTATION, THE IOL WAS EXPLANTED AND ANOTHER MODEL MULTIFOCAL IOL WAS IMPLANTED. THE PATIENT'S CURRENT OUTCOME IS UNKNOWN.

    Patient problems: Eye Pain; Inflammation; Toxic Anterior Segment Syndrome (TASS); Visual Impairment

    MAUDE report 21896496 ↗

Other Lens, Multifocal Intraocular devices