Skip to content

enVista Envy™ Toric IOL ETN (REF ETNU150+250)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Envy toric hydrophobic acrylic IOL (non-preloaded) +25.00D SE / +1.50D CYL · Class 3 · Implant

All models of enVista Envy Toric IOL (513)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770640956), received between Jul 15, 2025 and Jul 15, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Class IOpen, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Inflammation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT 1-DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE LEFT EYE (OS), THE PATIENT PRESENTED WITH MILD LOW-GRADE INFLAMMATION IN THE ANTERIOR CHAMBER. THE PATIENT NOTICED THE PROBLEM 1-WEEK TO 1-MONTH POST-OP AND DURING BOUTS OF INFLAMMATION, BEST CORRECT VISUAL ACUITY (BCVA) DECREASED BY 2 LINES. APPROXIMATELY 33 WEEKS AFTER IMPLANTATION, THE LENS WAS EXCHANGED FOR THE SAME MODEL & DIOPTER IOL. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Inflammation

    MAUDE report 22512628 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page