enVista Envy™ Toric IOL ETN (REF ETNU250+085)Recalls & reported problems
Bausch & Lomb Incorporated · enVista Envy toric hydrophobic acrylic IOL (non-preloaded) +8.50D SE / +2.50D CYL · Class 3 · Implant
All models of enVista Envy Toric IOL (513)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10757770641762), received between Aug 22, 2025 and Oct 16, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 27, 2025Class IOpen, Classified
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information2
Patient problems
- Endophthalmitis1
- Hypopyon1
- Toxic Anterior Segment Syndrome (TASS)1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE ISSUE RESOLVED AND VISUAL ACUITY OF AFFECTED EYE (RIGHT EYE) IS 20/40.
Patient problems: Endophthalmitis
MAUDE report 23309383 ↗InjuryAug 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT ONE DAY AFTER IMPLANTATION OF AN INTRAOCULAR LENS (IOL) INTO THE RIGHT EYE, THE PATIENT PRESENTED WITH 3+ CELL AND HYPOPYON. BASED ON THE FINDINGS THE SURGEON CONCLUDED THE CAUSE AS TOXIC ANTERIOR SEGMENT SYNDROME (TASS). IN THE SURGEON¿S OPINION, THE MOST LIKELY CAUSE OF THE EVENT MAY HAVE BEEN A CANNULA. PATIENT OUTCOME IS GOOD. TASS RESOLVED AFTER TREATMENT. THE FOLLOWING MEDICATIONS WERE PRESCRIBED (FOR USE AT HOME POST-OPERATIVELY): PMB (PRED/MOXI/BROM) 1%/0.5%/0.075% TID DUREZOL 0.05% EVERY HOUR MEDROL DOSE PACK. ON 19-AUG-25, RECEIVED AN UPDATE ON PATIENT'S OUTCOME AND VA FROM THE REPORTER. THE PATIENT VISITED TODAY (B)(6) 2025) AND TASS RESOLVED AFTER TREATMENT. S. SENG (B)(6) 2025.
Patient problems: Hypopyon; Toxic Anterior Segment Syndrome (TASS)
MAUDE report 22871708 ↗
Other Lens, Multifocal Intraocular devices
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular144 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular1 recall66 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular113 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular112 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular110 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular107 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular104 reports
- Clareon™ CNA0T0Alcon Laboratories, Inc. · Lens, multifocal intraocular93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.