Skip to content

enVista Envy™ Toric IOL ETN (REF ETNU250+225)Recalls & reported problems

Bausch & Lomb Incorporated · enVista Envy toric hydrophobic acrylic IOL (non-preloaded) +22.50D SE / +2.50D CYL · Class 3 · Implant

All models of enVista Envy Toric IOL (513)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10757770642042), received between Jan 16, 2026 and Jan 16, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 27, 2025Class IOpen, Classified

    In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

    Z-1680-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Blurred Vision1
  • Capsular Contracture1
  • Halo1
  • Inflammation1
  • Zonular Dehiscence1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 16, 2026Device Dislodged or Dislocated

    THIS EVENT WAS ORIGINALLY REPORTED AS EVENT #2 FOR PATIENT IDENTIFIER (A)(1) (B)(6). SINCE PATIENT IDENTIFIERS WERE LATER PROVIDED, THE CASE HAS BEEN REASSIGNED TO PATIENT IDENTIFIER (A)(1) (B)(6). ADDITIONAL INFORMATION WAS RECEIVED INDICATING SIX MONTHS AFTER IMPLANTATION OF THE IOL IN THE PATIENT'S LEFT EYE (OS), THE PATIENT REPORTED THAT VISION WAS NO LONGER AS CRISP AS THREE MONTHS PRIOR. EXAMINATION AT THAT TIME IDENTIFIED INFERIOR DECENTRATION OF THE IOL AND PHIMOSIS OF THE ANTERIOR CAPSULE. NINE MONTHS POST-IMPLANTATION, DURING THE LENS EXCHANGE, A CAPSULAR TENSION RING WAS IMPLANTED DUE TO ZONULAR DEHISCENCE. ACCORDING TO THE SURGEON, THE MOST LIKELY CAUSE OF THE EVENT WAS A CHRONIC INFLAMMATORY STIMULUS FROM THE LENS. FOLLOWING THE EXCHANGE, THE SURGEON DISCUSSED WITH THE PATIENT THE OPTIONS OF PRK OR LASIK TO ENHANCE DISTANCE VISION. THE PATIENT CONTINUES TO REPORT SEEING HALOS, WHICH THE SURGEON ATTRIBUTES TO THE PATIENT'S INABILITY TO ADAPT TO THE LENS DESIGN.

    Patient problems: Blurred Vision; Capsular Contracture; Halo; Inflammation; Zonular Dehiscence

    MAUDE report 24095015 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →