VITROS 6801059Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · VITROS ECI IMMUNODIAGNOSTIC SYSTEM - Refurbished · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10758750001040), received between Feb 17, 2026 and Feb 17, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 20, 2019Terminated

    Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems

    Z-1293-2019 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Low Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Low Test Results; Non Reproducible Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, LOWER THAN EXPECTED B-HCG RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE WHEN USING VITROS IMMUNODIAGNOSTICS PRODUCT B-HCG II REAGENT LOT 4570 ON A VITROS ECIQ SYSTEM. THE RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO REPEAT RESULTS OBTAINED FROM THE SAME PATIENT SAMPLE. PATIENT 1 VITROS B-HCG II RESULT OF <2.39 MIU/ML VS. THE EXPECTED RESULT OF 57.000 MIU/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER THAN EXPECTED VITROS B-HCG RESULT WAS NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT 613405.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24371204 ↗

Other FLUOROMETER, FOR CLINICAL USE devices

All FLUOROMETER, FOR CLINICAL USE devices →