VITROS 6802221Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Total B HCG II Calibrators · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10758750002337), received between Jan 7, 2025 and Jun 26, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER THREE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338741 ↗
  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER TWO OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338720 ↗
  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER ONE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338683 ↗
  • MalfunctionJan 7, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT A HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROL (LOT 85360) WHEN TESTED USING VITROS B-HCG II LOT 4280 ON A VITROS 5600 INTEGRATED SYSTEM. BIORAD LOT 85361 RESULT OF 11.091 MIU/ML VERSUS THE EXPECTED RESULT OF 5.908 MIU/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THE RESULTS WERE TO RECUR UNDETECTED. THE HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID; HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WERE NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THE EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093743 ↗

FDA clearance and classification

The FDA regulates product code JIT (calibrator, secondary) under 21 CFR 862.1150 as a Class II device. 21 CFR 862.1150 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VITROS 6802221

Has the VITROS 6802221 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10758750002337), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VITROS 6802221?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 7, 2025 and Jun 26, 2025. In those reports, the most common specific device problem is “High Test Results” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VITROS 6802221?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VITROS 6802221 FDA approved?

Devices of this type (product code JIT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Ortho Clinical Diagnostics.

Who makes the VITROS 6802221, and is it still sold?

The FDA's device database lists Ortho Clinical Diagnostics as the company responsible for it (brand VITROS, model 6802221). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.