VITROS 6802221Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Total B HCG II Calibrators · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10758750002337), received between Jan 7, 2025 and Jun 26, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER THREE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338741 ↗
  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER TWO OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338720 ↗
  • MalfunctionJun 26, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER ONE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338683 ↗
  • MalfunctionJan 7, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED THAT A HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROL (LOT 85360) WHEN TESTED USING VITROS B-HCG II LOT 4280 ON A VITROS 5600 INTEGRATED SYSTEM. BIORAD LOT 85361 RESULT OF 11.091 MIU/ML VERSUS THE EXPECTED RESULT OF 5.908 MIU/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THE RESULTS WERE TO RECUR UNDETECTED. THE HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID; HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WERE NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THE EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093743 ↗

Other CALIBRATOR, SECONDARY devices

All CALIBRATOR, SECONDARY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.