VITROS 6802221Recalls & reported problems
Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Total B HCG II Calibrators · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10758750002337), received between Jan 7, 2025 and Jun 26, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER THREE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338741 ↗MalfunctionJun 26, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER TWO OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338720 ↗MalfunctionJun 26, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED THAT HIGHER THAN EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROLS, LOT 1003900, WHEN USING VITROS BHCG REAGENT LOT 4600, ON A VITROS XT7600 SYSTEM. BIORAD L1 VITROS BHCG RESULTS OF 7.79, 7.54, 7.88, 8.12 AND 7.95 MIU/ML VS. THE BIORAD LOT 1003900 PACKAGE INSERT TARGET OF 4.59 MIU/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS BHCG RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THE CUSTOMER DID NOT PROCESS ANY PATIENT SAMPLES, SINCE THE BIORAD CONTROL RESULTS WERE OUTSIDE THE PI RANGES FOR THOSE FLUIDS. THERE WAS NO ALLEGATION OF HARM. THIS REPORT IS NUMBER ONE OF THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338683 ↗MalfunctionJan 7, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED THAT A HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LIQUICHEK IMMUNOASSAY PLUS CONTROL (LOT 85360) WHEN TESTED USING VITROS B-HCG II LOT 4280 ON A VITROS 5600 INTEGRATED SYSTEM. BIORAD LOT 85361 RESULT OF 11.091 MIU/ML VERSUS THE EXPECTED RESULT OF 5.908 MIU/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THE RESULTS WERE TO RECUR UNDETECTED. THE HIGHER-THAN-EXPECTED VITROS B-HCG II RESULT WAS OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID; HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLE RESULTS WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WERE NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THE EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21093743 ↗
Other CALIBRATOR, SECONDARY devices
- ARCHITECT 2K41-27Abbott Laboratories · CALIBRATOR, SECONDARY1 recall374 reports
- ARCHITECT 8K28-02Abbott Laboratories · CALIBRATOR, SECONDARY1 recall115 reports
- VITROS 8231474Ortho-Clinical Diagnostics, Inc. · Calibrator, secondary3 recalls0 reports
- BioPlex 2200 12000651Bio-Rad Laboratories, Inc. · Calibrator, secondary3 recalls0 reports
- ARCHITECT 08K2803Abbott GmbH · Calibrator, secondary1 recall93 reports
- ARCHITECT 7K63-25Abbott Ireland · CALIBRATOR, SECONDARY108 reports
- VITROS 8067324Ortho-Clinical Diagnostics, Inc. · Calibrator, secondary2 recalls0 reports
- VITROS 8568040Ortho-Clinical Diagnostics, Inc. · CALIBRATOR, SECONDARY1 recall18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.