Skip to content

ORTHO 715130Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · ANTI-FYA (ANTI-DUFFY A) 1X3ML

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10758750003365), received between Aug 12, 2024 and Aug 12, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2024False Positive Result; False Positive Result

    REPORT 2 OF 2 CMS (B)(4), WINDCHILL (B)(4) ON (B)(6) 2024, A CUSTOMER COMPLAINED TO AN ORTHO LABORATORY SPECIALIST ABOUT WHAT WAS DESCRIBED AS DISCORDANT POSITIVE FYA(FY1) TYPING RESULTS FOR TWO PATIENTS USING ORTHO ANTI-FYA(FY1) ANTI-SERUM LOT FA246A IN TUBE METHOD. COMPLAINANT: (B)(6), RED CELL REFERENCE LABORATORY MANAGER COMPLAINT REPORTER: (B)(6), ORTHO LABORATORY SPECIALIST SUPERVISOR EVENTS DATES: 10-12 JULY 2024; REPORTED ON (B)(6) 2024 BY (B)(6) WHO REPORTED IT TO ORTHO GLOBAL TECHNICAL SOLUTION CENTER (GTSC) ON THE SAME DAY. REAGENTS: ORTHO ANTI-FYA(FY1) ANTI-SERUM LOT FA246A, EXPIRY DATE 18 SEPTEMBER 2025, MANUFACTURED 18 SEPTEMBER 2023 COMPETITORS POLYSPECIFIC ANTI-HUMAN GLOBULIN LOT 030712101, EXPIRY DATE 23 FEBRUARY 2025 SOFTWARE VERSION: N/A PATIENT'S INFORMATION: EDTA SAMPLES. GENOTYPE CONFIRMED AS FY(A-B-); DAT NEGATIVE; FYA ANTIGEN NEGATIVE WITH OTHER SUPPLIERS OF ANTI-FYA(FY1) REAGENTS. PATIENTS ARE LIKELY (B)(6) ; ONE WAS V+VS+, THE OTHER WAS V-VS-. NO FURTHER DETAIL WAS PROVIDED. THE CUSTOMER REPORTED THAT BETWEEN (B)(6) 2024, THEY HAD TESTED 2 PATIENTS FOR FYA (FY1) ANTIGEN TYPING IN THE INDIRECT ANTIGLOBULIN TESTING (IAT) IN TUBE METHOD AND THAT THEY HAD OBTA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19962914 ↗
  • MalfunctionAug 12, 2024False Positive Result; False Positive Result

    REPORT 1 OF 2 CMS (B)(4), WINDCHILL (B)(4) ON 17 JULY 2024, A CUSTOMER COMPLAINED TO AN ORTHO LABORATORY SPECIALIST ABOUT WHAT WAS DESCRIBED AS DISCORDANT POSITIVE FYA(FY1) TYPING RESULTS FOR TWO PATIENTS USING ORTHO ANTI-FYA(FY1) ANTI-SERUM LOT FA246A IN TUBE METHOD. COMPLAINANT: (B)(6), RED CELL REFERENCE LABORATORY MANAGER COMPLAINT REPORTER: (B)(6), ORTHO LABORATORY SPECIALIST SUPERVISOR EVENTS DATES: 10-12 JULY 2024; REPORTED ON 17 JULY 2024 BY (B)(6) WHO REPORTED IT TO ORTHO GLOBAL TECHNICAL SOLUTION CENTER (GTSC) ON THE SAME DAY. REAGENTS: ORTHO ANTI-FYA (FY1) ANTI-SERUM LOT FA246A, EXPIRY DATE 18 SEPTEMBER 2025, MANUFACTURED 18 SEPTEMBER 2023 COMPETITOR POLYSPECIFIC ANTI-HUMAN GLOBULIN LOT 030712101, EXPIRY DATE 23 FEBRUARY 2025 SOFTWARE VERSION: N/A PATIENTS INFORMATION: EDTA SAMPLES. GENOTYPE CONFIRMED AS FY(A-B-); DAT NEGATIVES; FYA ANTIGEN NEGATIVE WITH OTHER SUPPLIERS OF ANTI-FYA(FY1) REAGENTS. PATIENTS ARE LIKELY AFRICAN; ONE WAS V+VS+, THE OTHER WAS V-VS-. NO FURTHER DETAIL WAS PROVIDED. THE CUSTOMER REPORTED THAT BETWEEN 10 AND 12 JULY 2024, THEY HAD TESTED 2 PATIENTS FOR FYA (FY1) ANTIGEN TYPING IN THE INDIRECT ANTIGLOBULIN TESTING (IAT) IN TUBE METHOD AND THAT THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19959983 ↗

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.