Skip to content

AFFIRMAGEN 719201Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · 0.8% Affirmagen® (Pooled Cells)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10758750003877), received between Jul 24, 2025 and Feb 9, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026False Negative Result

    (B)(4) / RA613355 COMPLAINT DESCRIPTION: ON (B)(6) 2026, A CUSTOMER CONTACTED ORTHO GLOBAL TECHNICAL SUPPORT CENTRE (GTSC) TO REPORT WHAT WAS DESCRIBED AS DISCREPANT NEGATIVE RESULTS WITH THE 0.8% AFFIRMAGEN A1 CELLS LOTS 8A052 AND 8A069 AND ORTHO MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 WHEN PERFORMING ABO TESTING FOR TWO PATIENT SAMPLES IN CONJUNCTION WITH THEIR ORTHO VISION ID-MTS ANALYZER (B)(6). COMPLAINANT/COMPLAINT REPORTER: (B)(6) - MEDICAL TECHNOLOGIST DATE OF EVENTS: 01 AND 14 JANUARY 2026 REPORTED ON: (B)(6) 2026 BY (B)(6) TO ORTHO GTSC REAGENTS: 0.8% AFFIRMAGEN LOT 8A052 EXPIRY DATE 03 FEBRUARY 2026 MANUFACTURE DATE 18 NOVEMBER 2025 0.8% AFFIRMAGEN LOT 8A069 EXPIRY DATE 03 MARCH 2026 MANUFACTURE DATE 16 DECEMBER 2025 MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 EXPIRY DATE 08 MAY 2026 MANUFACTURE DATE 08 AUGUST 2025 PATIENT INFORMATION: PATIENT ONE: SAMPLE ID (B)(6) PATIENT TWO: SAMPLE ID (B)(6) THE CUSTOMER REPORTED THAT THEY HAD TESTED SAMPLES FROM TWO PATIENTS FOR ABO FORWARD AND REVERSE GROUPING AND THAT THEY HAD OBTAINED THE FOLLOWING PATTERN OF REACTIONS: DATE OF TESTING PATIENT SAMPLE ID REAGENTS ANALYZER RESULTS (B)(6) 2

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24290468 ↗
  • MalfunctionFeb 9, 2026False Negative Result

    REPORT 2 OF 2 CMS (B)(6)/ RA613355 COMPLAINT DESCRIPTION: ON 14 JANUARY 2026, A CUSTOMER CONTACTED ORTHO GLOBAL TECHNICAL SUPPORT CENTRE (GTSC) TO REPORT WHAT WAS DESCRIBED AS DISCREPANT NEGATIVE RESULTS WITH THE 0.8% AFFIRMAGEN A1 CELLS LOTS 8A052 AND 8A069 AND ORTHO MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 WHEN PERFORMING ABO TESTING FOR TWO PATIENT SAMPLES IN CONJUNCTION WITH THEIR ORTHO VISION ID-MTS ANALYZER (B)(6). COMPLAINANT/COMPLAINT REPORTER: (B)(6) - MEDICAL TECHNOLOGIST DATE OF EVENTS: 01 AND 14 JANUARY 2026 REPORTED ON: 14 JANUARY 2026 BY (B)(6) TO ORTHO GTSC REAGENTS: 0.8% AFFIRMAGEN LOT 8A052 EXPIRY DATE 03 FEBRUARY 2026 MANUFACTURE DATE 18 NOVEMBER 2025 0.8% AFFIRMAGEN LOT 8A069 EXPIRY DATE 03 MARCH 2026 MANUFACTURE DATE 16 DECEMBER 2025 MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 EXPIRY DATE 08 MAY 2026 MANUFACTURE DATE 08 AUGUST 2025 PATIENT INFORMATION: PATIENT ONE: SAMPLE ID (B)(6) PATIENT TWO: SAMPLE ID (B)(6) THE CUSTOMER REPORTED THAT THEY HAD TESTED SAMPLES FROM TWO PATIENTS FOR ABO FORWARD AND REVERSE GROUPING AND THAT THEY HAD OBTAINED THE FOLLOWING PATTERN OF REACTIONS: DATE OF TESTING PATIENT SAMPLE ID REAGENTS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24290462 ↗
  • MalfunctionFeb 9, 2026False Negative Result

    REPORT 1 OF 2 (B)(6)/ (B)(4) COMPLAINT DESCRIPTION: ON (B)(6) 2026, A CUSTOMER CONTACTED ORTHO GLOBAL TECHNICAL SUPPORT CENTRE (GTSC) TO REPORT WHAT WAS DESCRIBED AS DISCREPANT NEGATIVE RESULTS WITH THE 0.8% AFFIRMAGEN A1 CELLS LOTS 8A052 AND 8A069 AND ORTHO MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 WHEN PERFORMING ABO TESTING FOR TWO PATIENT SAMPLES IN CONJUNCTION WITH THEIR ORTHO VISION ID-MTS ANALYZER (B)(6). COMPLAINANT/COMPLAINT REPORTER:(B)(6) - MEDICAL TECHNOLOGIST DATE OF EVENTS: 01 AND 14 JANUARY 2026 REPORTED ON: (B)(6) 2026 BY (B)(6) TO ORTHO GTSC REAGENTS: 0.8% AFFIRMAGEN LOT 8A052 EXPIRY DATE 03 FEBRUARY 2026 MANUFACTURE DATE 18 NOVEMBER 2025 0.8% AFFIRMAGEN LOT 8A069 EXPIRY DATE 03 MARCH 2026 MANUFACTURE DATE 16 DECEMBER 2025 MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT 070225037-09 EXPIRY DATE 08 MAY 2026 MANUFACTURE DATE 08 AUGUST 2025 PATIENT INFORMATION: PATIENT ONE: SAMPLE ID (B)(6) PATIENT TWO: SAMPLE ID (B)(6) THE CUSTOMER REPORTED THAT THEY HAD TESTED SAMPLES FROM TWO PATIENTS FOR ABO FORWARD AND REVERSE GROUPING AND THAT THEY HAD OBTAINED THE FOLLOWING PATTERN OF REACTIONS: DATE OF TESTING PATIENT SAMPLE ID REAGENTS ANALYZER RESULT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24290438 ↗
  • MalfunctionJul 24, 2025False Negative Result; False Negative Result

    CMS (B)(4) WINDCHILL RA611363 COMPLAINT DESCRIPTION: ON (B)(6) 2025, A CUSTOMER CONTACTED ORTHO GLOBAL TECHNICAL SUPPORT CENTER (GTSC) TO REPORT WHAT WAS DESCRIBED AS A FALSE NEGATIVE RESULT WITH THE 0.8% AFFIRMAGEN A1 CELLS LOT 8A943 WHEN PERFORMING ABO TESTING ON ONE PATIENT SAMPLE USING MTS ABD MONOCLONAL AND REVERSE GROUPING CARD LOT 022825037-10 IN CONJUNCTION WITH THEIR ORTHO VISION ID-MTS ANALYZER (B)(6). CUSTOMER NAME: (B)(6), MEDICAL TECHNOLOGIST (B)(6) CUSTOMER NO. (B)(6); (B)(6) HOSP, (B)(6) EVENT DATE: 12 JUNE 2025 AUTOMATED TESTING REAGENTS: MTS ABD MONOCLONAL AND REVERSE GROUPING CARD LOT 022825037-10 (EXPIRY DATE 20 DECEMBER 2025, MANUFACTURED DATE 20 MARCH 2025) 0.8% AFFIRMAGEN LOT 8A943 (EXPIRY 24 JUNE 2025, MANUFACTURED DATE 08 APRIL 2025) MTS DILUENT 2 PLUS LOT MDP251 (EXPIRY DATE 12 FEBRUARY 2026, MANUFACTURED DATE 12 FEBRUARY 2025). PATIENT INFORMATION: THE PATIENT WAS BORN IN (B)(6) 1955. SAMPLE ID IS (B)(6). THE PATIENT RECEIVED SCIG HOME INFUSIONS. THE PATIENT HAS NOT BEEN TRANSFUSED. NO FURTHER INFORMATION WAS PROVIDED. THE CUSTOMER REPORTED THAT ON (B)(6) 2025, ABO TESTING WAS PERFORMED ON ONE PATIENT SAMPLE USING MTS ABD MONOCLONAL AND REVERSE GROUPING CA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586312 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page