VITROS 8892382Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products CRBM Slides · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10758750005185), received between Nov 2, 2023 and May 22, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “High Test Results” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results6
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2025High Test Results; Low Test Results; High Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED RESULTS FROM MULTIPLE VITROS IMMUNORATE SLIDE ASSAYS WERE OBTAINED FROM TWO DIFFERENT LEVELS OF NON-VITROS THERMOFISHER MAS OMNICORE CONTROLS (MAS) LOT OCR2608 AND A SINGLE LEVEL OF VITROS THERAPEUTIC DRUG MONITORING PERFORMANCE VERIFIER (TDM PV) III LOT E2152 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS CARBAMAZEPINE (CRBM) SLIDE LOT 3920-0129-0168 MAS LOT OCR2608 L2 RESULTS OF 13.1 AND 12.6 UG/ML VS. THE EXPECTED RESULT OF 9.48 UG/ML. VITROS CRBM SLIDE LOT 3919-0128-9593 VITROS THERAPEUTIC DRUG MONITORING PERFORMANCE VERIFIER (TDM PV) III LOT E2152 RESULTS OF 9.0, 8.9, 6.3, 7.4, AND >20 UG/ML VS. THE EXPECTED RESULT OF 14.38 UG/ML. POST-SERVICE VITROS TDM PV III LOT E2152 RESULTS OF >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20, >20 AND >20 UG/ML VS. THE EXPECTED RESULT OF 14.38 UG/ML. VITROS PHENOBARBITAL (PHBR) SLIDE LOT 2516-0114-0757 MAS LOT OCR2608 L1 RESULTS OF 15.7, 17.1, 14.5, 15.7, 13.3, 15.0, 13.7, 13.5, 16.4 AND 15.8 UG/ML VS. THE EXPECTED RESULT OF 10.0 UG/ML. MAS LOT OCR2608 L2 RESULTS OF 26.0, 28.6, 2

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22070415 ↗
  • MalfunctionFeb 5, 2025High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS CARBAMAZEPINE (CRBM) RESULTS WERE OBTAINED FROM TWO LEVELS OF VITROS THERAPEUTIC DRUG MONITORING (TDM) PERFORMANCE VERIFIER (PV) QUALITY CONTROL (QC) MATERIAL USING VITROS CHEMISTRY PRODUCTS CRBM SLIDES LOT 3916-0125-4879 ON A VITROS XT 7600 INTEGRATED SYSTEM. EVENTS OBSERVED ON 19 DECEMBER 2024 AND 20 DECEMBER 2024: VITROS TDM PVI (LOT C2150) CRBM RESULT OF 9.20 AND 8.76 UG/ML VERSUS THE CUSTOMER'S BASELINE MEAN VALUE OF 4.83 UG/ML VITROS TDM PVIII (LOT E2152) CRBM RESULT OF 20.00 UG/ML VERSUS THE CUSTOMER'S BASELINE MEAN VALUE OF 14.89 UG/ML EVENTS OBSERVED ON 28 JANUARY 2025: VITROS TDM PVI (LOT C2150) CRBM RESULT OF 9.74 UG/ML VERSUS THE CUSTOMER'S BASELINE MEAN VALUE OF 4.83 UG/ML VITROS TDM PVIII (LOT E2152) CRBM RESULT OF 19.55 UG/ML VERSUS THE CUSTOMER'S BASELINE MEAN VALUE OF 14.89 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER-THAN-EXPECTED VITROS CRBM RESULTS WERE OBTAINED WHEN PROCESSING A NON-PATIENT QUALITY CONTROL FLUID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21317890 ↗
  • MalfunctionMar 7, 2024High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A HIGHER THAN EXPECTED VITROS CRBM RESULT WAS OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD MULTIQUAL UNASSAYED CONTROLS, LOT 56720 TESTED ON A VITROS 5600 INTEGRATED SYSTEM. BIORAD LOT 56720 LEVEL 1 CRBM RESULT OF 7.42 UG/ML VS. THE EXPECTED RESULT OF 3.75 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS CRBM RESULT WAS OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID, HOWEVER, THE INVESTIGATION COULD NOT RULE OUT THAT PATIENT SAMPLES WOULD BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WAS NO REPORTED ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4) .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855595 ↗
  • MalfunctionNov 2, 2023High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS CARBAMAZEPINE (CRBM) RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD IMMUNOASSAY PLUS CONTROL LOT 85310 TESTED ON A XT7600 INTEGRATED SYSTEM. BIORAD LOT 85310 LEVEL 2 CRBM RESULTS OF 9.69, 9.24 AND 9.79 UG/ML VS. THE EXPECTED RESULT OF 6.22 UG/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS CRBM RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THERE WAS NO REPORTED ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT 603682.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18059275 ↗
  • MalfunctionNov 2, 2023High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS CARBAMAZEPINE (CRBM) RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD IMMUNOASSAY PLUS CONTROL LOT 85310 TESTED ON A XT7600 INTEGRATED SYSTEM. BIORAD LOT 85310 LEVEL 2 CRBM RESULTS OF 9.69, 9.24 AND 9.79 UG/ML VS. THE EXPECTED RESULT OF 6.22 UG/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS CRBM RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THERE WAS NO REPORTED ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18059250 ↗
  • MalfunctionNov 2, 2023High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS CARBAMAZEPINE (CRBM) RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD IMMUNOASSAY PLUS CONTROL LOT 85310 TESTED ON A XT7600 INTEGRATED SYSTEM. BIORAD LOT 85310 LEVEL 2 CRBM RESULTS OF 9.69, 9.24 AND 9.79 UG/ML VS. THE EXPECTED RESULT OF 6.22 UG/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER THAN EXPECTED VITROS CRBM RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THERE WAS NO REPORTED ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF THREE MDR¿S FOR THIS EVENT. THREE 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THREE DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18058631 ↗

Other ENZYME IMMUNOASSAY, CARBAMAZEPINE devices

All ENZYME IMMUNOASSAY, CARBAMAZEPINE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.