1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Sep 12, 2023Class IIOpen, Classified

    Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.

    Z-0236-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code CHI (lipase-esterase, enzymatic, photometric, lipase) under 21 CFR 862.1465 as a Class I device. 21 CFR 862.1465 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VITROS 1668409

Has the VITROS 1668409 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0236-2024 (Class II), began on Sep 12, 2023: Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the VITROS 1668409 FDA approved?

Devices of this type (product code CHI) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Ortho-Clinical Diagnostics, Inc.

Who makes the VITROS 1668409, and is it still sold?

The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 1668409). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE devices

All LIPASE-ESTERASE, ENZYMATIC, PHOTOMETRIC, LIPASE devices →