VITROS 6197735Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products BRAHMS PCT Reagent Pack · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10758750033287), received between Aug 25, 2023 and Mar 10, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2
  • High Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PROCALCITONIN (PCT) RESULT WAS OBTAINED FROM A SAMPLE FROM A SINGLE PATIENT WHEN TESTED ON A VITROS XT 7600 INTEGRATED SYSTEM USING REAGENT LOT 0440 WHEN COMPARED TO A REPEAT PCT TEST RESULT OBTAINED FROM A REPEAT TEST OF THE SAME SAMPLE ON THE SAME VITROS XT 7600 INTEGRATED SYSTEM. PATIENT SAMPLE RESULT OF 1.916 NG/ML VERSUS THE EXPECTED (REPORTED) REPEAT RESULT OF 0.059 NG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED, THE NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PCT RESULT OF 1.916 NG/ML WAS QUESTIONED BY THE LABORATORY STAFF AND REPEAT TEST EVENTS WERE SUBSEQUENTLY PROCESSED. THE QUESTIONED RESULT WAS NOT REPORTED FROM THE LABORATORY AND THERE HAS BEEN NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21559949 ↗
  • MalfunctionDec 22, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS B¿R¿A¿H¿M¿S PROCALCITONIN (PCT) QC FLUID RESULTS WERE OBTAINED FROM VITROS B¿R¿A¿H¿M¿S PCT LOT 0360 LEVEL 2 QUALITY CONTROL (QC) FLUID, USING VITROS IMMUNODIAGNOSTIC PRODUCTS B¿R¿A¿H¿M¿S PCT REAGENT ON A VITROS 5600 INTEGRATED SYSTEM VITROS B¿R¿A¿H¿M¿S PCT LEVEL 2 RESULTS OF 0.03 AND 0.03 NG/ML VERSUS THE EXPECTED RESULT OF 1.91 NG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER THAN EXPECTED VITROS B¿R¿A¿H¿M¿S PCT RESULTS WERE FROM NON-PATIENT FLUID. THE CUSTOMER DID NOT GIVE ANY INDICATION THAT PATIENT RESULTS HAD BEEN AFFECTED AND THERE IS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18382573 ↗
  • MalfunctionAug 25, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS IMMUNODIAGNOSTICS PRODUCTS B¿R¿A¿H¿M¿S PCT REAGENT PACK (PCT) RESULTS WERE OBTAINED FROM 10 DIFFERENT PATIENT SAMPLES AND 2 DIFFERENT LEVELS OF VITROS PCT LOT 0320 CONTROLS TESTED ON A VITROS 3600 IMMUNODIAGNOSTIC SYSTEM. PATIENT SAMPLE 1 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 8.76 NG/ML PATIENT SAMPLE 2 VITROS PCT RESULT OF 0.226 NG/ML VS. THE EXPECTED RESULT OF 2.25 NG/ML PATIENT SAMPLE 3 VITROS PCT RESULT OF 0.062 NG/ML VS. THE EXPECTED RESULT OF 9.28 NG/ML PATIENT SAMPLE 5 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 5.59 NG/ML PATIENT SAMPLE 6 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 1.22 NG/ML PATIENT SAMPLE 7 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 3.14 NG/ML PATIENT SAMPLE 9 VITROS PCT RESULT OF 0.037 NG/ML VS. THE EXPECTED RESULT OF 4.31 NG/ML PATIENT SAMPLE 12 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 3.77 NG/ML PATIENT SAMPLE 13 VITROS PCT RESULT OF <0.030 NG/ML VS. THE EXPECTED RESULT OF 12.8 NG/ML PATIENT SAMPLE 14 VITROS PCT RESULT OF 0.100 NG/M

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17627054 ↗

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.