Skip to content

Bard® Fluoro-4™ Silicone Ureteral Stent 233624Recalls & reported problems

C. R. Bard, Inc. · Bard® Fluoro-4™ Silicone Ureteral Stent with Hydro-Glide™ Guidewire · Class 2 · Implant

All models of Bard Fluoro-4 Silicone Ureteral Stent (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10801741011679), received between Sep 6, 2017 and Apr 22, 2022 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break8
  • Component Missing1
  • Device Damaged Prior to Use1
  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Inaccurate Flow Rate1
  • Incorrect Measurement1
  • Material Fragmentation1

Patient problems

  • No Known Impact Or Consequence To Patient5
  • No Clinical Signs, Symptoms or Conditions4
  • No Patient Involvement3
  • No Information2
  • Urinary Retention1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHEN DOCTOR WANTS TO DISASSEMBLE THE STENT, IT WAS FOUND THAT THE D-J WAS BROKEN. IT WAS STATED THAT THEY REPLACED IT WITH A NEW ONE TO CONTINUE THE OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14189809 ↗
  • MalfunctionApr 1, 2022Migration; Migration

    IT WAS REPORTED THAT THE USER EXPERIENCED INCOMPLETE EMPTYING, HEMATURIA, STENT MIGRATION AND DISLODGMENT WHILE USING THE URETERAL STENTS. MEDICAL INTERVENTION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13980834 ↗
  • MalfunctionApr 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE USER EXPERIENCED INCOMPLETE EMPTYING, HEMATURIA, STENT MIGRATION AND DISLODGMENT WHILE USING URETERAL STENTS. MEDICAL INTERVENTION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13980437 ↗
  • MalfunctionApr 1, 2022Inaccurate Flow Rate; Inaccurate Flow Rate

    IT WAS REPORTED THAT THE USER EXPERIENCED INCOMPLETE EMPTYING, HEMATURIA, STENT MIGRATION AND DISLODGMENT WHILE USING URETERAL STENTS. MEDICAL INTERVENTION WAS UNKNOWN.

    Patient problems: Urinary Retention

    MAUDE report 13980220 ↗
  • MalfunctionDec 26, 2021Component Missing; Component Missing

    IT WAS REPORTED THAT THERE WAS NO URETERAL STENT IN THE INNER PACKAGING. THE CUSTOMER USED ANOTHER NEW URETERAL STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13085711 ↗
  • MalfunctionSep 25, 2019Partial Blockage; Break; Partial Blockage

    IT WAS REPORTED THAT THE END TIP OF THE STENT WAS EMBRITTLED AND BROKEN PRIOR TO USE.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 9117645 ↗
  • MalfunctionMay 9, 2019Break; Device Damaged Prior to Use

    IT WAS REPORTED THAT THE GUIDE WIRE WAS FOUND BROKEN UPON OPENING THE PACKAGE.

    Patient problems: No Patient Involvement

    MAUDE report 8596596 ↗
  • MalfunctionSep 10, 2018Break; Incorrect Measurement; Failure to Advance; Incorrect Measurement

    IT WAS REPORTED THAT THE STENT SEEMED TOO NARROW AND CAUSED THE GUIDEWIRE NOT TO PASS THROUGH, SUBSEQUENTLY THE STENT BROKE WHILE TRYING TO REMOVE IT FROM THE GUIDEWIRE.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 7862000 ↗
  • MalfunctionSep 7, 2017Break

    IT WAS REPORTED THAT THE STENT WAS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6847711 ↗
  • MalfunctionSep 6, 2017Break

    IT WAS REPORTED THAT THE STENT WAS BROKEN.

    Patient problems: No Information

    MAUDE report 6846496 ↗
  • MalfunctionSep 6, 2017Break

    IT WAS REPORTED THAT THE STENT WAS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6846435 ↗
  • MalfunctionSep 6, 2017Break

    IT WAS REPORTED THAT THE STENT WAS BROKEN. THE AFFECTED DEVICE HAS EITHER ONE OF THE LOT #, NGBN 2439 OR NGAW 4423, AS PROVIDED BY THE HEAD OR NURSE.

    Patient problems: No Information

    MAUDE report 6846302 ↗
  • MalfunctionSep 6, 2017Break

    IT WAS REPORTED THAT THE STENT WAS BROKEN. THE AFFECTED DEVICE HAS EITHER ONE OF THE LOT #'S, NGBN2439 OR NGAW4423, AS PROVIDED BY THE HEAD OR NURSE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6846299 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page