Skip to content

Bard® InLay Optima® Multilength Ureteral Stent with NiCore™ Guidewire 786600Recalls & reported problems

C. R. Bard, Inc. · INLAY OPTIMA MULTI-LENGTH URETERAL STENT W/NITINOL GUIDEWIRE · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10801741015073), received between Mar 28, 2018 and Jul 25, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem3
  • Incorrect Measurement2
  • Patient Device Interaction Problem2
  • Fracture1
  • Knotted1
  • Material Fragmentation1
  • Migration1

Patient problems

  • Discomfort5
  • Pain5
  • No Clinical Signs, Symptoms or Conditions2
  • Hematuria1
  • Injury1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Unspecified Kidney or Urinary Problem1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT BEFORE USE, STENT WAS ALREADY BROKEN IN THE PACKAGE. THEY SAW THAT THE CURL WAS BROKEN OFF OF THE STENT. SO, IT COULD NOT BE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22608729 ↗
  • InjuryFeb 19, 2025Migration; Migration

    IT WAS REPORTED THAT ON (B)(6) 2025, OPTIMA WAS INSERTED IN AN EMERGENCY FOR CALCULOUS PYELONEPHRITIS. AFTER THE NEXT DAY, THE PATIENT EXPERIENCED HEMATURIA AND BACK PAIN, AND CT SCANS WERE TAKEN, AND IT SEEMS THAT THE TIP OF THE STENT WAS STRAYED INTO THE RENAL CORTEX. THE BACK PAIN WOULD NOT BE AFFECTED BY STENT. ON (B)(6) 2025, THE DOCTOR WAS CONTACTED AND TOLD THAT THE PATIENT HAD A COAGRA IN THE RENAL CORTEX AND HAD FEVER, SO CORRECTED OR REPLACED THE POSITION OF STENT FROM THEN.

    Patient problems: Hematuria; Pain

    MAUDE report 21410418 ↗
  • MalfunctionDec 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE AFTER OPENING THE PRODUCT, THERE WERE CRACKS NEAR THE URETHRAL STENT PIGTAIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21036792 ↗
  • MalfunctionMar 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE "786600 - INLAY OPTIMATM MULTI-LENGTH URETERAL STENT WITH NICORE¿ GUIDEWIRE, 6 FR., 22CM ¿ 32CM " AND THE PATIENT EXPERIENCED "PAIN OR DISCOMFORT" AND "URINARY SYMPTOMS". THIS DID NOT RESULTED IN THE PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.

    Patient problems: Discomfort; Pain; Unspecified Kidney or Urinary Problem

    MAUDE report 18931490 ↗
  • MalfunctionMar 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    CRITICAL CARE PHYSICIAN REPORTED IN A PMCF THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE "787624 - INLAY OPTIMATM URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE, 6.0 FR. X 24CM " AND THE PATIENT EXPERIENCED "PAIN OR DISCOMFORT". THIS DID NOT RESULT IN THE PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.

    Patient problems: Discomfort; Pain

    MAUDE report 18913982 ↗
  • MalfunctionMar 19, 2019Incorrect Measurement; Material Integrity Problem; Incorrect Measurement

    IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL, AND AS A RESULT IT DID NOT CURL INTO THE PATIENT'S RENAL PELVIS; ALLEGEDLY THIS CAUSED A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THIS EVENT HAS OCCURRED WITH THE LAST 3 PATIENTS. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON (B)(6) 2019 FROM IBC REPRESENTATIVE, ALL THREE PATIENTS ALLEGEDLY EXPERIENCED DISCOMFORT DUE TO THE LONGER STENT.

    Patient problems: Discomfort

    MAUDE report 8431860 ↗
  • MalfunctionMar 18, 2019Material Integrity Problem

    IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL. AS A RESULT IT DID NOT CURL NICELY INTO THE PATIENT'S RENAL PELVIS, ALLEGEDLY CAUSING A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THAT THE EVENT HAPPENED WITH THE LAST 3 PATIENTS.

    Patient problems: Discomfort; No Consequences Or Impact To Patient; Pain

    MAUDE report 8429671 ↗
  • MalfunctionMar 18, 2019Incorrect Measurement; Material Integrity Problem; Incorrect Measurement

    IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL. AS A RESULT, IT DID NOT CURL NICELY INTO THE PATIENT'S RENAL PELVIS, ALLEGEDLY CAUSING A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THAT THE EVENT HAD HAPPENED WITH THE LAST 3 PATIENTS.

    Patient problems: Discomfort; Pain

    MAUDE report 8429669 ↗
  • InjuryMar 28, 2018Knotted

    IT WAS REPORTED THAT THE STENT WAS INSERTED INTO THE PATIENT; HOWEVER, THE STENT HAD A KNOT MID URETER. DR. (B)(6) SAW THE PATIENT AND DISCOVERED THE FAULTY STENT. THE PATIENT WAS TAKEN TO SURGERY TO REMOVE THE STENT. A LAPAROSCOPIC PYELOPLASTY AND PYELOLITHOTOLY WAS PERFORMED TO REMOVE THE STENT.

    Patient problems: Injury; No Code Available

    MAUDE report 7378397 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page