Bard® InLay Optima® Multilength Ureteral Stent with NiCore™ Guidewire 786600Recalls & reported problems
C. R. Bard, Inc. · INLAY OPTIMA MULTI-LENGTH URETERAL STENT W/NITINOL GUIDEWIRE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10801741015073), received between Mar 28, 2018 and Jul 25, 2025 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem3
- Incorrect Measurement2
- Patient Device Interaction Problem2
- Fracture1
- Knotted1
- Material Fragmentation1
- Migration1
Patient problems
- Discomfort5
- Pain5
- No Clinical Signs, Symptoms or Conditions2
- Hematuria1
- Injury1
- No Code Available1
- No Consequences Or Impact To Patient1
- Unspecified Kidney or Urinary Problem1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT BEFORE USE, STENT WAS ALREADY BROKEN IN THE PACKAGE. THEY SAW THAT THE CURL WAS BROKEN OFF OF THE STENT. SO, IT COULD NOT BE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608729 ↗InjuryFeb 19, 2025Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2025, OPTIMA WAS INSERTED IN AN EMERGENCY FOR CALCULOUS PYELONEPHRITIS. AFTER THE NEXT DAY, THE PATIENT EXPERIENCED HEMATURIA AND BACK PAIN, AND CT SCANS WERE TAKEN, AND IT SEEMS THAT THE TIP OF THE STENT WAS STRAYED INTO THE RENAL CORTEX. THE BACK PAIN WOULD NOT BE AFFECTED BY STENT. ON (B)(6) 2025, THE DOCTOR WAS CONTACTED AND TOLD THAT THE PATIENT HAD A COAGRA IN THE RENAL CORTEX AND HAD FEVER, SO CORRECTED OR REPLACED THE POSITION OF STENT FROM THEN.
Patient problems: Hematuria; Pain
MAUDE report 21410418 ↗MalfunctionDec 30, 2024Fracture; Fracture
IT WAS REPORTED THAT THE AFTER OPENING THE PRODUCT, THERE WERE CRACKS NEAR THE URETHRAL STENT PIGTAIL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21036792 ↗MalfunctionMar 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE "786600 - INLAY OPTIMATM MULTI-LENGTH URETERAL STENT WITH NICORE¿ GUIDEWIRE, 6 FR., 22CM ¿ 32CM " AND THE PATIENT EXPERIENCED "PAIN OR DISCOMFORT" AND "URINARY SYMPTOMS". THIS DID NOT RESULTED IN THE PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.
Patient problems: Discomfort; Pain; Unspecified Kidney or Urinary Problem
MAUDE report 18931490 ↗MalfunctionMar 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
CRITICAL CARE PHYSICIAN REPORTED IN A PMCF THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE "787624 - INLAY OPTIMATM URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE, 6.0 FR. X 24CM " AND THE PATIENT EXPERIENCED "PAIN OR DISCOMFORT". THIS DID NOT RESULT IN THE PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.
Patient problems: Discomfort; Pain
MAUDE report 18913982 ↗MalfunctionMar 19, 2019Incorrect Measurement; Material Integrity Problem; Incorrect Measurement
IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL, AND AS A RESULT IT DID NOT CURL INTO THE PATIENT'S RENAL PELVIS; ALLEGEDLY THIS CAUSED A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THIS EVENT HAS OCCURRED WITH THE LAST 3 PATIENTS. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON (B)(6) 2019 FROM IBC REPRESENTATIVE, ALL THREE PATIENTS ALLEGEDLY EXPERIENCED DISCOMFORT DUE TO THE LONGER STENT.
Patient problems: Discomfort
MAUDE report 8431860 ↗MalfunctionMar 18, 2019Material Integrity Problem
IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL. AS A RESULT IT DID NOT CURL NICELY INTO THE PATIENT'S RENAL PELVIS, ALLEGEDLY CAUSING A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THAT THE EVENT HAPPENED WITH THE LAST 3 PATIENTS.
Patient problems: Discomfort; No Consequences Or Impact To Patient; Pain
MAUDE report 8429671 ↗MalfunctionMar 18, 2019Incorrect Measurement; Material Integrity Problem; Incorrect Measurement
IT WAS REPORTED THAT THE STENT WAS LONGER THAN USUAL. AS A RESULT, IT DID NOT CURL NICELY INTO THE PATIENT'S RENAL PELVIS, ALLEGEDLY CAUSING A LOT OF DISCOMFORT AND PAIN. IT WAS ALSO REPORTED THAT THE STENT KNOTTED DUE TO THE INCREASED LENGTH. THE PHYSICIAN REPORTED THAT THE EVENT HAD HAPPENED WITH THE LAST 3 PATIENTS.
Patient problems: Discomfort; Pain
MAUDE report 8429669 ↗InjuryMar 28, 2018Knotted
IT WAS REPORTED THAT THE STENT WAS INSERTED INTO THE PATIENT; HOWEVER, THE STENT HAD A KNOT MID URETER. DR. (B)(6) SAW THE PATIENT AND DISCOVERED THE FAULTY STENT. THE PATIENT WAS TAKEN TO SURGERY TO REMOVE THE STENT. A LAPAROSCOPIC PYELOPLASTY AND PYELOLITHOTOLY WAS PERFORMED TO REMOVE THE STENT.
Patient problems: Injury; No Code Available
MAUDE report 7378397 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page