Skip to content

ARROW IPN000096 (REF AI-07155-KS)Recalls & reported problems

Teleflex Incorporated · Temporary Pacing Catheter with Shrouded Pins/Introducer Kit · Class 2

All models of ARROW (1488)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10801902004533), received between Oct 25, 2023 and Sep 6, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K945229); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak7
  • Material Integrity Problem2
  • Capturing Problem1
  • Difficult to Open or Close1
  • Intermittent Capture1
  • Leak/Splash1
  • Loose or Intermittent Connection1
  • Migration1

Patient problems

  • Insufficient Information7
  • No Clinical Signs, Symptoms or Conditions3
  • Arrhythmia1
  • Bacteremia1
  • Hemorrhage/Blood Loss/Bleeding1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 6, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50638971 ↗
  • MalfunctionSep 6, 2026Fluid/Blood Leak; Patient Device Interaction Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombosis/Thrombus

    MAUDE report 50638822 ↗
  • MalfunctionSep 6, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50638676 ↗
  • MalfunctionAug 25, 2026Intermittent Capture; Fluid/Blood Leak; Loose or Intermittent Connection; Migration

    Patient problems: Arrhythmia; Insufficient Information

    MAUDE report 50422578 ↗
  • InjuryJul 30, 2026Fluid/Blood Leak

    It was reported that in the CVICU, nursing staff observed intravenous fluid and blood leaking into the Cath-Gard contamination shield while the temporary transvenous pacing (TVP) system was in use. The TVP had been in place for approximately 24 hours when the issue was identified and was subsequently removed on (b)(6)2026. The patient experienced an unknown amount of blood loss; however, no blood transfusion or administration of blood products was reported. It was further reported that the patient became febrile and was subsequently diagnosed with a central line-associated bloodstream infection (CLABSI). The patient received intravenous antibiotic treatment from (b)(6)2026 through (b)(6)2026. The patient subsequently underwent permanent pacemaker placement(PPM) and was discharged home on (b)(6)2026, with a subsequent discharge reported on (b)(6)2026. No additional medical or surgical interventions were reported beyond those described above.

    Patient problems: Bacteremia; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50088596 ↗
  • MalfunctionFeb 24, 2026Material Integrity Problem

    IT WAS REPORTED " THE BIGGEST PROBLEM IS THE TUOHY THAT IS SUPPOSED TO TIGHTEN ONTO THE PACING CATHETER DOES NOT GET TIGHT ENOUGH TO KEEP THE CATHETER IN PLACE, WHICH ALLOWS THE CATHETER TO MOVE WITH MINIMAL EFFORT. BLOOD BACKUP INTO THE SWANDOM ALMOST IMMEDIATELY AFTER PLACEMENT AND CONTINUED AFTER CHECKING THE TUOHY WAS AS TIGHT AS POSSIBLE. ONCE A PRESSURE BAG WAS HOOKED UP TO THE ONE PATIENT, THE ENTIRE SWANDOM FILLED WITH FLUID (IT WAS DESCRIBED AS LOOKING LIKE A LONG WATER BALLOON). THE PATIENT DID FINE UNTIL THE NEXT DAY; WE GOT LUCKY THAT THE CATHETER DIDN'T MOVE ENOUGH THAT IT STOPPED CAPTURING".

    Patient problems: Insufficient Information

    MAUDE report 24429354 ↗
  • MalfunctionFeb 23, 2026Material Integrity Problem

    IT WAS REPORTED " THE LOCKING MECHANISM, EVEN WHEN FULLY TIGHTENED, DOES NOT SECURE THE PACING WIRE IN PLACE. WE'VE ALSO NOTICED THAT BLOOD/FLUID BACKS UP INTO THE SWANDOM ALMOST IMMEDIATELY AS WELL. WE HAD A PATIENT YESTERDAY, WHO BY THE TIME THEY GOT UP TO THEIR ROOM, WAS NOT LONGER CAPTURING. THAT PATIENT HAD TO BE BROUGHT BACK TO OUR LAB AND WE WE'RE STILL UNABLE TO SECURE THE WIRE IN PLACE. LUCKILY OUR EP PHYSICIAN WAS FREE TO PLACE A PERMANENT PACEMAKER AT THIS TIME, BUT THIS COULD HAVE EASILY BEEN A MAJOR PROBLEM HAD THE PATIENT NOT HAD A DECENT UNDERLYING RATE. "

    Patient problems: Insufficient Information

    MAUDE report 24415632 ↗
  • MalfunctionDec 8, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT "SIDE INFUSION PART WON'T GIVE BLOOD RETURN. IT SUCKS IN LARGE AMOUNTS OF AIR FORM THE SITE OF THE PACEMAKER WIRE ". ADDITIONAL INFORMATION STATES THAT ANOTHER CATHETER WAS USED TO COMPLETE THE PROCEDURE. AT THE TIME OF THIS REPORT, NO ADDITIONAL INFORMATION HAS BEEN RECEIVED FROM THE CUSTOMER.

    Patient problems: Insufficient Information

    MAUDE report 23742075 ↗
  • MalfunctionMay 8, 2024Fluid/Blood Leak; Difficult to Open or Close; Fluid/Blood Leak; Difficult to Open or Close

    IT WAS REPORTED THAT "SOMETHING HAPPEN THAT CAUSED THE VALVE NOT TO CLOSE AFTER THE LOCKING AND CONTAMINATION SLEEVE WERE ATTACHED. THE SHEATH WAS BLEEDING BACK INTO THE SLEEVE AND STAFF NEEDED TO CHANGE OUT THE WHOLE SYSTEM. UPON REMOVING THE SHEATH THEY DID NOT NOTICE ANY DAMAGE OR LEAKS".

    Patient problems: Insufficient Information

    MAUDE report 19271892 ↗
  • MalfunctionNov 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE REPORT STATES, "BLOOD RETURN FROM TEMP WIRE AREA WITHIN THE SWANDOM AND BLOOD WITHIN THE SHEATH. I FOLLOWED UP WITH CC TEAM AND MENTIONED PATIENT WAS SHIPPED TO BI BOSTON FOR FURTHER EP TESTING WITH TEMP WIRE IN PLACE. I WAS TOLD DR DID SEE PATIENT OVER THE WEEKEND, BLEEDING WAS MAINTAINED. TEMP WIRE INDICATION- BRADYCARDIA, SYNCOPE".

    Patient problems: Insufficient Information

    MAUDE report 18133665 ↗
  • MalfunctionOct 25, 2023Capturing Problem; Capturing Problem

    IT WAS REPORTED THAT WHILE IN USE, THE PACER WIRE IS NOT CONDUCTING. AT THE TIME OF THIS REPORT THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.

    Patient problems: Insufficient Information

    MAUDE report 18003910 ↗

FDA clearance and classification

Cleared through 510(k) K945229 on Aug 8, 1995 (decision: Substantially Equivalent). Submitted by Arrow Intl., Inc. as “Arrow Multipolar Intracardiac Electrode Catheters”. The FDA review took 286 days. FDA record ↗

The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K945229. One submission often covers many models, so check the product and lot details in each record.

  • Z-3197-2018 · Jul 30, 2018 · Terminated
    ARROW Bipolar Electrode Catheter with shrouded pins , 6 Fr, 110 cm Product Code: AI-06210-S Product Usage: Arrow electrophysiology catheters are intended for temporary use in elec

Common questions about the ARROW IPN000096

Has the ARROW IPN000096 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10801902004533), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the ARROW IPN000096?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Oct 25, 2023 and Sep 6, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (7), “Material Integrity Problem” (2) and “Capturing Problem” (1); the most common patient problems are “Arrhythmia” (1), “Bacteremia” (1) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ARROW IPN000096?

None of the reports are classified as deaths, 1 as an injury and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ARROW IPN000096 FDA approved?

It was cleared through 510(k) K945229 on Aug 8, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ARROW IPN000096, and is it still sold?

The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN000096). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page