Skip to content

ARROW IPN030769Recalls & reported problems

Teleflex Incorporated · Pediatric Multi-Lumen Central Venous Catheterization Set with Blue FlexTip(R) Catheter · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10801902027464), received between Aug 29, 2023 and Oct 28, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

24 other recalls cover devices cleared under the same FDA submission (K862056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Fluid/Blood Leak1
  • Material Twisted/Bent1
  • Mechanical Jam1
  • Physical Resistance/Sticking1
  • Unraveled Material1

Patient problems

  • Hematoma1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • Needle Stick/Puncture1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Difficult to Advance; Material Twisted/Bent; Physical Resistance/Sticking; Difficult to Advance; Material Twisted/Bent; Physical Resistance/Sticking

    IT WAS REPORTED THAT "ON (B)(6) 2025, IN PAEDIATRIC INTENSIVE CARE, DURING THE INSERTION OF A CENTRAL VENOUS CATHETER, THE GUIDE DID NOT ADVANCE PROPERLY AND IT BECAME STUCK. WHEN THE GUIDE WAS REMOVED, IT LOOKED TWISTED AND BENT. THE CATHETER COULD NOT BE INSERTED. THE INCIDENT CAUSED SLIGHT BLEEDING AT THE PUNCTURE SITE. A SECOND CATHETER WAS INSERTED, REQUIRING A SECOND PUNCTURE." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Needle Stick/Puncture

    MAUDE report 23401922 ↗
  • MalfunctionOct 16, 2024Unraveled Material; Mechanical Jam; Unraveled Material; Mechanical Jam

    IT WAS REPORTED "DURING INSERTION OF THE CVC, THE SWG GOT STUCK." A NEW CVC WAS INSERTED. THERE WAS A HEMATOMA ON THE INSERTION SIGHT. THE DEVICE WAS REMOVED AND REPLACED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: Hematoma

    MAUDE report 20460904 ↗
  • MalfunctionAug 29, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE CVC WAS INSTALLED IN A PEDIATRIC PATIENT AND "ON DAY 4 OF THE INSTALLATION, HE PRESENTED A LEAK IN ONE OF THE LUMENS". "DURING THE HOSPITALISATION, ON (B)(6) 2023, A LARGE LEAK WAS OBSERVED , APPARENTLY AT THE INSERTION SITE OF THE CENTRAL VENOUS CATHETER (CVC) LOCATED IN THE RIGHT SUBCLAVIAN ARTERY INSTALLED ON (B)(6) 2023. THE STATE OF THE 3 LUMENS WAS CHECKED BY INFUSING PHYSIOLOGICAL SALINE, WITH THE PROXIMAL AND DISTAL LUMEN REFLUXING AND INFUSING WITHOUT DIFFICULTY , WHILE THE MEDIAL LUMEN SHOWED AN ABUNDANT OUTFLOW OF SALINE. SEDATION IS ADMINISTERED TO THE PATIENT FOR PUNCTURE AND INSTALLATION OF A 13-CM, 3-LUMEN 4 FR CENTRAL VENOUS CATHETER (CVC), WHICH IS FIXED WITH FIXATION POINTS AT THE INSERTION SITE, ONE IN BOTH ORIFICES OF THE CVC FIXATION SYSTEM, AT 10 CMS. THE 3 LUMENS REFLUX AND INFUSE WITHOUT DIFFICULTY."

    Patient problems: Insufficient Information

    MAUDE report 17646105 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K862056. One submission often covers many models, so check the product and lot details in each record.

  • Z-2377-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE
  • Z-1119-2019 · Jan 10, 2019 · Terminated
    Arrow Two-Lumen Central Venous Catheterization Set with Blue FlexTip Catheter Product Code: CD-10802 GTN: 00801902015396
  • Z-1118-2019 · Jan 10, 2019 · Terminated
    Arrow Two-Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter Product Code: AH-11802 (OUS) GTN: 00801902026132
  • Z-0723-2019 · Nov 14, 2018 · Terminated
    Arrow CVC 2 Lumen, Pediatric Two-Lumen Central Venous Cauterization Set with Blue FlexTip Catheter, 4 Fr 2 Lumen 5cm, Reference # CS-12402 The Arrow CVC is intended to provide sho
  • Z-2819-2018 · Jul 10, 2018 · Terminated
    HANDS-OFF Multi-Lumen Central Venous Catheter with Blue FlexTip for use only with Arrow One-Piece Percutaneous Sheath Introducer System, Material HO-14703 The central venous cath
  • Z-2818-2018 · Jul 10, 2018 · Terminated
    HANDS-OFF Multi-Lumen Central Venous Catheter with Blue FlexTip for use only with Arrow One-Piece Percutaneous Sheath Introducer System, Material HO-14702 The central venous cathe
  • Z-2336-2018 · May 11, 2018 · Terminated
    Multi-Lumen Central Venous Catheterization Set with Blue FlexTip Catheter, CS-15853-E Product CVCs are intended to permit venous access to the central circulation through a single
  • Z-1978-2018 · Apr 11, 2018 · Terminated
    Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) Catheter and Sharps Safety Features; Product Code: ASK-15402-HOP
  • Z-1967-2018 · Apr 11, 2018 · Terminated
    Maximal Barrier Drape; Product Code: ASK-04000-HMC
  • Z-1982-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue(R) MAC(TM) Two-Lumen Central Venous Access Kit for use with 7-7.5 Fr. Catheters; Product Code: ASK-21142-WHC2
  • Z-1332-2018 · Dec 22, 2017 · Terminated
    (1) Access Tray, Part Number (PN): ASK-04001-MS; Arrow Multi-Lumen CVC Kit (2) PN: CDC-12703-1A; (3) PN: CDC-15703-1A; (4) PN: CDC-15703-B1A; Arrow Two-Lumen CVC Kit (5) PN: CDC-12
  • Z-0648-2018 · Oct 17, 2017 · Terminated
    Multi Lumen Central Venous Catheterization Kit with Blue FlexTip Catheter and Sharps Safety Features; Product Code: ASK 15703 NYP; Exp. Dates Feb 2018 - Nov 2018
  • Z-0657-2018 · Oct 17, 2017 · Terminated
    Central Venous Catheterization Kit with Blue FlexTip Catheter; Product code: AK-04301; Exp. Dates Feb 2018 - Nov 2018
  • Z-0654-2018 · Oct 17, 2017 · Terminated
    Arrow Multi Lumen CVC Kit (Central venous catheter); Product Code: CDC-15703-1A; Exp. Dates Feb 2018 - Nov 2018
  • Z-0650-2018 · Oct 17, 2017 · Terminated
    Multi Lumen Central Venous Catheterization Kit; Product Code: ASK-12703-HH2; Exp. Dates Feb 2018 - Nov 2018

…and 9 more.

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page