ARROW IPN036440 (REF CDC-42703-XP1A)Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features, and Maximal Barrier Preca · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10801902117080), received between Sep 25, 2023 and Nov 12, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
54 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation3
- Material Twisted/Bent2
- Connection Problem1
- Unraveled Material1
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryNov 12, 2025Material Separation; Material Separation
IT WAS REPORTED "THE CATHETER BROKE OFF FROM THE BASE/TIP WITH NO RESISTANCE AND COULD NOT BE RETRIEVED UNDER THE PATIENT'S SKIN." THE CATHETER WAS REMOVED BY IR RETRIEVAL. THERE WAS NO SIGNS OR SYMPTOMS AFTER THE EVENT. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE AND DISCHARGED HOME".
Patient problems: Foreign Body In Patient
MAUDE report 23532416 ↗MalfunctionApr 24, 2024Connection Problem; Connection Problem
IT WAS REPORTED "THE INTRODUCER NEEDLE SLIDES OFF THE SYRINGE IF GEL OR BLOOD GETS ON IT. THE INTRODUCER NEEDLE DOES NOT STAY STATIONARY WHEN PLACE ON THE SYRINGE". ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER, NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 19172911 ↗MalfunctionMar 13, 2024Unraveled Material; Material Twisted/Bent; Unraveled Material; Material Twisted/Bent
IT WAS REPORTED "SPRING WIRE GUIDE KINKS EASILY". NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2024-00226, 9680794-2024-00227, AND 9680794-2024-00225.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18893813 ↗MalfunctionMar 13, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "SPRING WIRE GUIDE KINKS EASILY". NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2024-00226, 9680794-2024-00310, 9680794-2024-00227.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18893723 ↗MalfunctionSep 25, 2023Material Separation; Material Separation
MEDWATCH REPORT RECEIVED: "ON (B)(6) 2023 THIS PATIENT HAD A 20CM LEFT SUBCLAVIAN CVC PLACED BY IR. IT FUNCTIONED PROPERLY, UNTIL (B)(6) 2023 WHEN RN HEARD A LITTLE POP AND SAW THAT THE DISTAL LUMEN HAD COME OFF AT THE PLASTIC BROWN PART AND BLOOD WAS LEAKING OUT. RN CLAMPED LINE AND TIED A KNOT IN THE LUMEN. MD CONTACTED AND CENTRAL LINE WAS REMOVED. REF REPORT: MW5120619." NO PATIENT HARM WAS REPORTED. THE CVC WAS REMOVED AND A NEW ONE PLACED. THE PATIENT'S CONDITION IS REPORTED AS "STABLE".
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17810220 ↗MalfunctionSep 25, 2023Material Separation; Material Separation
MEDWATCH REPORT RECEIVED: "ON (B)(6) 2023 THIS PATIENT HAD A 20CM LEFT SUBCLAVIAN CVC PLACED BY IR. IT FUNCTIONED PROPERLY, UNTIL (B)(6) 2023 WHEN RN HEARD A LITTLE POP AND SAW THAT THE DISTAL LUMEN HAD COME OFF AT THE PLASTIC BROWN PART AND BLOOD WAS LEAKING OUT. RN CLAMPED LINE AND TIED A KNOT IN THE LUMEN. MD CONTACTED AND CENTRAL LINE WAS REMOVED. REF REPORT: MW5120620." NO PATIENT HARM WAS REPORTED. THE CVC WAS REMOVED AND A NEW ONE PLACED. THE PATIENT'S CONDITION IS REPORTED AS "STABLE".
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17809843 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.
- Z-2376-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
- Z-1914-2026 · Mar 11, 2026 · Open, ClassifiedARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
- Z-2123-2026 · Jan 7, 2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
- Z-0003-2024 · Aug 10, 2023 · Open, ClassifiedArrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0002-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0351-2023 · Oct 27, 2022 · TerminatedArrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
- Z-0352-2023 · Oct 27, 2022 · TerminatedPressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
- Z-1604-2022 · May 25, 2022 · Open, ClassifiedTELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
- Z-1995-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip, ARROWg+ard Blue PLUS Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product
- Z-1990-2018 · Apr 11, 2018 · TerminatedPressure Injectable Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Code: ASK-45802-PHMW
- Z-2000-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Code: ASK-42703-SJH
- Z-1996-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; P
- Z-1988-2018 · Apr 11, 2018 · TerminatedPressure Injectable ARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Codes: ASK-42703-PAM1, ASK-42703-PSHR1, ASK-45703-PNY, ASK-45703-PWHC1
- Z-1994-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS Pressure Injectable Multi-Lumen CVC Kit; Product Codes: CDC-45703-P1A; CDC-45703-PB1A; CDC-42703-P1A; and CDC-15703-P1A
- Z-1980-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit; Product Code: ASK-15703-PHMC
…and 39 more.
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page