ARROW IPN001366Recalls & reported problems

Teleflex Incorporated · Arrow-Clark(TM) VectorFlow(TM) Retrograde Hemodialysis Catheterization Set Long-Term Access. Catheter Length, Tip to Cuff: 27 cm · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10801902127355), received between Sep 11, 2023 and Feb 3, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Incomplete or Inadequate Connection” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K141051); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incomplete or Inadequate Connection7
  • Material Split, Cut or Torn5
  • Leak/Splash2
  • Adverse Event Without Identified Device or Use Problem1
  • Air/Gas in Device1
  • Crack1
  • Expiration Date Error1
  • Fitting Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Insufficient Information1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Incomplete or Inadequate Connection

    IT WAS REPORTED THAT "THE CUSTOMER COMPLAINS THAT THE CATHETERS ARE NO LONGER SEALED AFTER THE SEALING RING HAS BEEN OVERWRITTEN AND THE CATHETERS HAVE BEEN SCREWED TOGETHER. IT HAS BEEN FOUND THAT THE MATERIAL OF THE SEALING RING IS PROBABLY TOO BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT CLEARLY SHOW THAT THE SEALING RINGS ARE TORN". NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251193 ↗
  • MalfunctionFeb 3, 2026Incomplete or Inadequate Connection

    IT WAS REPORTED THAT "THE CUSTOMER COMPLAINS THAT THE CATHETERS ARE NO LONGER SEALED AFTER THE SEALING RING HAS BEEN OVERWRITTEN AND THE CATHETERS HAVE BEEN SCREWED TOGETHER. IT HAS BEEN FOUND THAT THE MATERIAL OF THE SEALING RING IS PROBABLY TOO BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251175 ↗
  • MalfunctionDec 11, 2025Material Split, Cut or Torn; Incomplete or Inadequate Connection; Material Split, Cut or Torn; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT: "THE CUSTOMER COMPLAINS THAT THE CATHETERS DONT SEALED WHEN SCREWED TOGETHER. THEY FEEL THAT THE MATERIAL OF THE SEALING RING IS BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT SHOW THAT THE RINGS ARE TORN. THERE WAS NO HARM TO THE PATIENTS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23779838 ↗
  • MalfunctionDec 11, 2025Material Split, Cut or Torn; Incomplete or Inadequate Connection; Material Split, Cut or Torn; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT: "THE CUSTOMER COMPLAINS THAT THE CATHETERS DONT SEALED WHEN SCREWED TOGETHER. THEY FEEL THAT THE MATERIAL OF THE SEALING RING IS BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT SHOW THAT THE RINGS ARE TORN. THERE WAS NO HARM TO THE PATIENTS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777604 ↗
  • MalfunctionDec 11, 2025Material Split, Cut or Torn; Incomplete or Inadequate Connection; Material Split, Cut or Torn; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT: "THE CUSTOMER COMPLAINS THAT THE CATHETERS DONT SEALED WHEN SCREWED TOGETHER. THEY FEEL THAT THE MATERIAL OF THE SEALING RING IS BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT SHOW THAT THE RINGS ARE TORN. THERE WAS NO HARM TO THE PATIENTS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777491 ↗
  • MalfunctionDec 11, 2025Material Split, Cut or Torn; Incomplete or Inadequate Connection; Material Split, Cut or Torn; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT: "THE CUSTOMER COMPLAINS THAT THE CATHETERS DONT SEALED WHEN SCREWED TOGETHER. THEY FEEL THAT THE MATERIAL OF THE SEALING RING IS BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT SHOW THAT THE RINGS ARE TORN. THERE WAS NO HARM TO THE PATIENTS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23776913 ↗
  • MalfunctionDec 11, 2025Material Split, Cut or Torn; Incomplete or Inadequate Connection; Material Split, Cut or Torn; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT: "THE CUSTOMER COMPLAINS THAT THE CATHETERS DONT SEAL WHEN SCREWED TOGETHER. THEY FEEL THAT THE MATERIAL OF THE SEALING RING IS BRITTLE AND IS DAMAGED WHEN THE CATHETER IS SCREWED TOGETHER. I HAVE TWO EXAMPLES THAT SHOW THAT THE RINGS ARE TORN. THERE WAS NO HARM TO THE PATIENTS."

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23776870 ↗
  • MalfunctionFeb 12, 2025Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT "WHEN THE PHYSICIAN TRIED TO SCREW THE THREADED COMPRESSION CAP ONTO THE HUB ASSEMBLY IT WOULD NOT TIGHTEN. THE CAP TURNS AND TURNS AND DOESN'T SECURE ONTO HUB". A NEW SET WAS OPENED TO RETRIEVE THE HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21365422 ↗
  • MalfunctionJan 12, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT: DURING DIALYSIS, AIR WAS DETECTED IN THE ASPIRATION AND BLOOD LEAKAGE FROM THE LINE DURING INFUSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18501014 ↗
  • MalfunctionNov 8, 2023Expiration Date Error; Adverse Event Without Identified Device or Use Problem; Expiration Date Error; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT: ON THE (B)(6) 2023 A CATHETER WAS INSERTED SUCCESSFULLY INTO A PATIENT. THEY REALISED AFTER INSERTION THAT IT WAS 10 DAYS EXPIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18092422 ↗
  • MalfunctionSep 20, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT: "FOLLOWING CATHETER OVER GUIDE WIRE REPLACEMENT, PATENCY OF THE CATHETER WAS CONFIRMED WITH CONTRAST UNDER VISION BY THE DOCTOR, HOWEVER, SAYS HE COULDN'T GET BLOOD BACK AS HE ASPIRATED. INSTEAD, HE WAS PULLING AIR BUT WHEN HE FLUSH WITH HEPARIN FLUSHING WAS FLOWING EASILY. THE DOCTOR REMOVED THE CATHETER." THE DEVICE WAS REMOVED AND ANOTHER BRAND OF CATHETER WAS INSERTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17785962 ↗
  • MalfunctionSep 11, 2023Leak/Splash; Air/Gas in Device; Leak/Splash; Air/Gas in Device

    IT WAS REPORTED THAT: "NON-OCCLUSIVE RING AND UNSCREWED OLIVE. AIR BUBBLE IN CIRCUIT AND CATHETER." THE CONNECTION WAS LEAKING AIR. THE CATHETER WAS INSERTED IN THE LEFT INTERNAL JUGULAR VEIN. NO PATIENT HARM WAS REPORTED. THE DIALYSIS PROCEDURE WAS DELAYED AND THE CATHETER WAS REMOVED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17722452 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141051. One submission often covers many models, so check the product and lot details in each record.

  • Z-2157-2026 · Apr 10, 2026 · Open, Classified
    Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →