ARROW IPN038265Recalls & reported problems
Teleflex Incorporated · RediGuard(R) IAB · Class 2
All models of ARROW (1494)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10801902144208), received between Mar 31, 2022 and Sep 16, 2026 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (8 reports).
2 other recalls cover devices cleared under the same FDA submission (K981660); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem8
- Device Alarm System2
- Material Integrity Problem2
- Complete Blockage1
- Gas/Air Leak1
- Leak/Splash1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information5
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788318 ↗MalfunctionSep 2, 2026Inflation Problem
Patient problems: Insufficient Information
MAUDE report 50566118 ↗MalfunctionOct 21, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED "BLOOD IN HELIUM TUBING". ADDITIONAL INFOMATION STATES THAT THE IAB CATHETER WAS REMOVED AND IT IS UNKNOWN IF IT WAS REPLACED. THE CUSTOMER HAS NOT PROVIDED ADDITIONAL INFORMATION RELATED TO THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 23347167 ↗MalfunctionAug 20, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED "CCL - FAILURE TO UNWRAP. FAILURE TO UNWRAP WHILE PLACING". ADDITIONAL INFORMATION STATES THAT MANUAL INFLATION WAS ATTEMPTED AND UNSUCCESSFUL IN AIDING THE IAB CATHETER TO UNWRAP". ANOTHER IAB CATHETER WAS INSERTED. THE LOCATION OF THE SECOND DEVICE IS UNKNOWN. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 22849560 ↗InjuryJun 3, 2025Device Alarm System; Device Alarm System
IT WAS REPORTED "PERSISTENT POSSIBLE HELIUM LOSS 3 ALARMS". ADDITIONAL INFORMATION STATES THE IAB CATHETER WAS REMOVED AND REPLACED. THE SECOND IAB CATHETER WAS INSERTED INTO A DIFFERENT INSERTION SITE. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22134092 ↗MalfunctionMay 27, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED "BALLOON WAS NOT FULLY INFLATED AT THE TIP". THE USER REMOVED THE BALLOON AND INSERTED ANOTHER. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22090886 ↗MalfunctionMar 19, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED "REPORTED FAILURE TO UNWRAP.HAD IMMEDIATELY RECEIVED HIGH PRESSURE ALARMS. THERE WERE NO VISIBLE KINKS IN THE LINE AND TEE VERIFIED APPROPRIATE PLACEMENT. MANUAL INFLATION NOT ATTEMPTED.INITIAL IAB REMOVED AND REPLACED WITH NO FURTHER ALARMS RECEIVED. IAB VOLUME DECREASED INCREMENTALLY TO 45CC TO MAINTAIN PLATEAU PRESSURE WITHIN 25MMHG OF AUG. PATIENT REPORTED TO HAVE REMAINED STABLE THROUGHOUT". NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21644434 ↗MalfunctionSep 13, 2024Material Rupture; Material Rupture
IT WAS REPORTED "RUPTURED BALLOON". ADDITIONAL INFORMATION STATES TO CONTINUE THERAPY, A SECOND CATHETER WAS ISERTED. THE SECOND IAB CATHETER WAS INSERTED INTO THE SAME INSERTION SITE. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20217189 ↗MalfunctionSep 13, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED "CCL PLACED IAB AND STARTED RECEIVING HELIUM LOSS ALARMS, DECIDED TO REMOVE THE CATHETER. UPON REMOVAL [THE USER] OBSERVED 2 KINKS IN THE SHEATH. THEY DID NOT PLACE A NEW IAB. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20216717 ↗MalfunctionDec 21, 2023Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE PUMP WAS ALARMING FOR "CATHETER ALARMS" AFTER THE PATIENT RETURNED FROM THE OR. THE PATIENT IS CRITICAL, VENTILATED, HAS ATRIAL FIBRILLATION WITH EPISODES OF V-TACH ON DOBUTAMINE AND EPINEPHRINE AT HIGH DOSES. THE USER WAS UNABLE TO ASPIRATE FROM THE CENTRAL LUMEN. AS A RESULT, THE IAB WAS REMOVED AND A 2ND IAB WAS INSERTED AT THE SAME INSERTION SITE. NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18372706 ↗MalfunctionNov 2, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT "IAB PLACEMENT ATTEMPTED ON LF, NO ISSUES OR DIFFICULTY NOTED WITH INSERTION, CONSTANT HIGH PRESSURE ALARMS ON PUMP, PUMP EXCHANGED, HIGH PRESSURE ALARMS CONTINUED. VOLUME DECREASED TO 40CC. MANUAL INFLATION NOT ATTEMPTED DUE TO LACK OF VISUAL PLACEMENT (IN OR) CXR CONFIRMED PROPER PLACEMENT. 2ND IAB INSERTED ON RF WITH SAME RESULT. 2ND IAB WAS KEPT FOR RETURN PHOTOS SHOW IAB STILL TIGHTLY WRAPPED. [THE PERFUSIONIST] STATED HE ALSO ATTEMPTED TO INFLATE IAB WITH THE PUMP IN INTERNAL TRIGGER AFTER THE IAB WAS REMOVED. HIGH PRESSURE ALARMS STILL NOTED AND IAB DID NOT UNWRAP. PATIENT WAS MEDICALLY STABILIZED AND TRANSFERRED TO THE ICU WHERE HE REMAINED STABLE WITHOUT IAB PLACEMENT". SEE ASSOCIATED MDR #3010532612-2023-00628.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18057917 ↗MalfunctionSep 7, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT "AFTER INSERTION AND CONNECTION TO THE PUMP, A HELIUM LEAKAGE WAS NOTICED, AND THE CARDIOLOGIST DECIDED TO REMOVE THIS IAB AND PLACE A NEW ONE. NO HELIUM LEAKAGE WAS FOUND, AND THE THERAPY WAS CONTINUED". THE 2ND IAB WAS INSERTED AT THE SAME INSERTION SITE. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17701132 ↗MalfunctionJun 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE "BALLOON WOULD NOT FULLY INFLATE". AS A RESULT, THE CATHETER WAS REMOVED AND A 2ND IAB WAS INSERTED. AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED. ASSOCIATED MDR #3010532612-2023-00320.
Patient problems: Insufficient Information
MAUDE report 17100753 ↗MalfunctionJun 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE "BALLOON WOULD NOT FULLY INFLATE". AS A RESULT, THE CATHETER WAS REMOVED AND A 2ND IAB WAS INSERTED. AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED. SEE ASSOCIATED MDR #3010532612-2023-00319.
Patient problems: Insufficient Information
MAUDE report 17099225 ↗MalfunctionJun 9, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED VIA A TWITTER POST THAT THERE WERE ISSUES ENCOUNTERED WITH AN "'ARROW IAB". AFTER FOLLOW UP WITH CUSTOMER IT WAS REPORTED THAT A 50CC BALLOON CATHETER WAS NOT INFLATING. AS A RESULT, A 2ND IAB WAS INSERTED AT THE SAME INSERTION SITE. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "STABLE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17099222 ↗MalfunctionMar 31, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED BY THE CLINICAL SUPPORT SPECIALIST (CSS) THAT THE IAB WAS INSERTED WITHOUT DIFFICULTY OR INCIDENT, WHEN THE STAFF ATTEMPTED TO INITIATE PUMPING, HIGH PRESSURE ALARMS OCCURRED X2. THERE WERE NO NOTED KINKS AND THE IAB APPEARED TO NOT UNWRAP UNDER FLUOROSCOPY. THE MD OPTED TO EXCHANGE THE IAB. A SECOND IAB WAS INSERTED IN THE SAME SITE. THERE WAS MINIMAL DELAY AND THE PATIENT REMAINED STABLE THROUGHOUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13967116 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K981660. One submission often covers many models, so check the product and lot details in each record.
- Z-1061-2016 · Feb 10, 2016 · TerminatedRediGuard IAB: 7Fr 30cc; Product Code: IAB-S730C The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto
- Z-1063-2016 · Feb 10, 2016 · TerminatedPercutaneous Insertion Tray; Product Code: IAK-S7IT The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during dia
Other System, balloon, intra-aortic and control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page