Skip to content

ARROW IPN039779Recalls & reported problems

Teleflex Incorporated · RediGuard(R) IAB · Class 2

All models of ARROW (1494)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10801902145632), received between Oct 11, 2022 and Apr 17, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Device Alarm System1
  • Material Integrity Problem1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Device Alarm System

    IT WAS REPORTED "AFTER THE CATHETER INSERTED INTO THE PATIENT, THE PUMP TRIGGERED A HELIUM LEAKAGE ALARM. CHANGE THE CATHETER". NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".

    Patient problems: Insufficient Information

    MAUDE report 24912414 ↗
  • MalfunctionApr 16, 2026Material Integrity Problem

    IT WAS REPORTED "THERE WAS A LOSS OF BARRIER FUNCTION WHEN THE CONDOM MET THE INTRODUCER". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Insufficient Information

    MAUDE report 24896268 ↗
  • MalfunctionNov 17, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE PATIENT UNDERWENT CAG, PTCA WITH POSTOPERATIVE IABP SUPPORT, THE SURGEON INSERTED THE BALLOON, WHEN THE BALLOON WAS WITHDRAWN FROM THE BOX THERE WAS A SLIGHT KINKED OF THE BALLOON, [THE SURGEON] TRIED TO INSERT THE BALLOON, THE INSERT PROCESS WAS NOT SMOOTH AND COULD NOT BE INSERTED, THE BALLOON WAS WITHDRAWN AND REPLACED WITH A NEW BALLOON, THE IMPLANTATION WAS SUCCESSFUL". FURTHER INFORMATION STATES THAT THE SECOND CATHETER WAS INSERTED AT THE SAME INSERTION SITE. THE PATIENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15815029 ↗
  • MalfunctionOct 27, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT "THE PATIENT WAS ADMITTED TO (B)(6) ON (B)(6) 2022 AND WAS D IAGNOSED WITH 'ACUTE INFERIOR WALL MYOCARDIAL INFARCTION'. AFTER DISCUSSION IN THE DEPARTMENT, 'EMERGENCY CAG + PCI IF NECESSARY' WAS PROPOSED. INTRAOPERATIVELY, THE PATIENT HAD POOR CARDIAC FUNCTION AND WAS TREATED WITH IABP SUPPORT. AFTER THE BALLOON WAS INSERTED, THE PUMP ALARMED, AND THEN BLOOD WAS FOUND IN THE BALLOON, WHICH WAS JUDGED TO BE A BALLOON RUPTURE. THE BALLOON WAS REPLACED WITH A NEW ONE AND WAS INSERTED SUCCESSFULLY, WITH NORMAL COUNTERPULSATION WORKING". THE SECOND CATHETER WAS INSERTED AT THE SAME INSERTION SITE. THE CURRENT PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15683611 ↗
  • MalfunctionOct 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT "PATIENT SUFFERED FROM ACUTE MYOCARDIAL INFARCTION AND WAS SENT TO THE CCU FOR EMERGENCY TREATMENT. AFTER THE OPERATION, THE IAB CATHETER WAS INSERTED FOR ASSISTANCE. BLOOD WAS FOUND, AFTER CHECKING, IT WAS FOUND THAT THE BALLOON WAS RUPTURED. AS A RESULT, THE CATHETER WAS REPLACED AT THE SAME INSERTION SITE AND WORKED AS EXPECTED. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY OR DEATH."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15575223 ↗

Other System, balloon, intra-aortic and control devices

All System, balloon, intra-aortic and control devices →