Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10801902145632), received between Oct 11, 2022 and Apr 17, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Device Alarm System1
- Material Integrity Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Device Alarm System
IT WAS REPORTED "AFTER THE CATHETER INSERTED INTO THE PATIENT, THE PUMP TRIGGERED A HELIUM LEAKAGE ALARM. CHANGE THE CATHETER". NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".
Patient problems: Insufficient Information
MAUDE report 24912414 ↗MalfunctionApr 16, 2026Material Integrity Problem
IT WAS REPORTED "THERE WAS A LOSS OF BARRIER FUNCTION WHEN THE CONDOM MET THE INTRODUCER". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 24896268 ↗MalfunctionNov 17, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT "THE PATIENT UNDERWENT CAG, PTCA WITH POSTOPERATIVE IABP SUPPORT, THE SURGEON INSERTED THE BALLOON, WHEN THE BALLOON WAS WITHDRAWN FROM THE BOX THERE WAS A SLIGHT KINKED OF THE BALLOON, [THE SURGEON] TRIED TO INSERT THE BALLOON, THE INSERT PROCESS WAS NOT SMOOTH AND COULD NOT BE INSERTED, THE BALLOON WAS WITHDRAWN AND REPLACED WITH A NEW BALLOON, THE IMPLANTATION WAS SUCCESSFUL". FURTHER INFORMATION STATES THAT THE SECOND CATHETER WAS INSERTED AT THE SAME INSERTION SITE. THE PATIENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15815029 ↗MalfunctionOct 27, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "THE PATIENT WAS ADMITTED TO (B)(6) ON (B)(6) 2022 AND WAS D IAGNOSED WITH 'ACUTE INFERIOR WALL MYOCARDIAL INFARCTION'. AFTER DISCUSSION IN THE DEPARTMENT, 'EMERGENCY CAG + PCI IF NECESSARY' WAS PROPOSED. INTRAOPERATIVELY, THE PATIENT HAD POOR CARDIAC FUNCTION AND WAS TREATED WITH IABP SUPPORT. AFTER THE BALLOON WAS INSERTED, THE PUMP ALARMED, AND THEN BLOOD WAS FOUND IN THE BALLOON, WHICH WAS JUDGED TO BE A BALLOON RUPTURE. THE BALLOON WAS REPLACED WITH A NEW ONE AND WAS INSERTED SUCCESSFULLY, WITH NORMAL COUNTERPULSATION WORKING". THE SECOND CATHETER WAS INSERTED AT THE SAME INSERTION SITE. THE CURRENT PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15683611 ↗MalfunctionOct 11, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "PATIENT SUFFERED FROM ACUTE MYOCARDIAL INFARCTION AND WAS SENT TO THE CCU FOR EMERGENCY TREATMENT. AFTER THE OPERATION, THE IAB CATHETER WAS INSERTED FOR ASSISTANCE. BLOOD WAS FOUND, AFTER CHECKING, IT WAS FOUND THAT THE BALLOON WAS RUPTURED. AS A RESULT, THE CATHETER WAS REPLACED AT THE SAME INSERTION SITE AND WORKED AS EXPECTED. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY OR DEATH."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15575223 ↗
Other System, balloon, intra-aortic and control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page