ARROW IPN918400Recalls & reported problems
Teleflex Incorporated · Arrowg+ard Blue(R) MAC(TM) Two-Lumen Central Venous Access for use with 7.5 - 8 Fr. Catheters · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10801902179286), received between Aug 8, 2023 and Apr 17, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
10 other recalls cover devices cleared under the same FDA submission (K011761); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Gas/Air Leak2
- Biocompatibility1
- Break1
- Complete Blockage1
- Device Appears to Trigger Rejection1
- Failure to Advance1
- Physical Resistance/Sticking1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions3
- Low Blood Pressure/ Hypotension2
- Anaphylactic Shock1
- Angioedema1
- Extravasation1
- Hypersensitivity/Allergic reaction1
- Renal Impairment1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Failure to Advance; Physical Resistance/Sticking
IT WAS REPORTED THAT: "UNABLE TO FLOAT SWAN CATHETER THROUGH THE MAC. A NEW KIT WAS OPENED FOR THE SWAN SHEATH TO BE REPLACED. THERE WAS NO REPORTED PATIENT HARM." ASSOCIATED COMPLAINT 9680794-2026-00322.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24910180 ↗MalfunctionDec 26, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "THE MAC WAS INDWELLING WITHIN THE PATIENT WHEN THE NURSE PULLED BACK ON THE LUMEN AND HAD NO BLOOD RETURN. WHEN THE NURSE DID GET BLOOD RETURN, BUBBLES WERE OBSERVED. ONE HOUR LATER THE NURSE NOTICED WETNESS ON THE PATIENT'S SKIN. THE NURSE FLUSHED THE CATHETER AND NOTICED SWELLING WITHIN THE PATIENT'S NECK. THE DECISION WAS MADE TO REMOVE THE CATHETER DUE TO CONCERN OVER POSSIBLE INFILTRATION AND/OR CATHETER DAMAGE". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: Extravasation
MAUDE report 23896318 ↗MalfunctionNov 19, 2025Complete Blockage; Complete Blockage
IT WAS REPORTED THAT "MAC INSERTED AND USED IN CVOR SURGERY THEN PATIENT TRANSPORTED TO CVICU. CVICU REPORTED CLOT IN DISTAL (BROWN) LUMEN WITHIN 30 MINUTES OF ARRIVAL AND CANNOT ASPIRATE THROUGH WHITE PROXIMAL LUMEN". ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO PATIENT HARM OR INJURY. A NEW CATHETER WAS NOT INSERTED. THE PATIENT STATUS IS REPORTED AS "FINE'.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23596344 ↗MalfunctionOct 15, 2025Break; Break
IT WAS REPORTED THAT "WE HAVE A KIT THAT HAD A BROKEN COMPONENT IF IT WHEN OPENED." THIS WAS FOUND PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT. THE USER CHANGED TO A NEW DEVICE TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23278496 ↗MalfunctionAug 9, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE CENTRAL LINE WAS PLACED IN THE OR BY ANESTHESIA ON (B)(6) 2024. WHILE TRYING TO DRAW LABS ON (B)(6) 2024, THE RN NOTICED THAT WHILE TRYING TO PULL BACK FROM THE INTRODUCER PORT, ALL SHE COULD DRAW BACK WAS AIR. THE LINE WAS INSPECTED AND USING A 10 ML SYRINGE APPROXIMATELY 100 ML OF AIR WAS REMOVED. ALL OF THE OTHER LUMENS PULLED BACK BLOOD WITH NO ISSUE. THE PATIENT HAD NO SUBCUTANEOUS AIR PALPABLE". ASSOCIATED COMPLAINTS (B)(4).
Patient problems: Insufficient Information
MAUDE report 19948500 ↗MalfunctionAug 9, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "THE CENTRAL LINE WAS PLACED IN THE OR BY ANESTHESIA ON (B)(6) 2024. WHILE TRYING TO DRAW LABS ON (B)(6) 2024, THE RN NOTICED THAT WHILE TRYING TO PULL BACK FROM THE INTRODUCER PORT, ALL SHE COULD DRAW BACK WAS AIR. THE LINE WAS INSPECTED AND USING A 10 ML SYRINGE APPROXIMATELY 100 ML OF AIR WAS REMOVED. ALL OF THE OTHER LUMENS PULLED BACK BLOOD WITH NO ISSUE. THE PATIENT HAD NO SUBCUTANEOUS AIR PALPABLE". ASSOCIATED COMPLAINTS 9680794-2024-00755, 9680794-2024-00754 AND 9680794-2024-00753
Patient problems: Insufficient Information
MAUDE report 19948490 ↗MalfunctionAug 9, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT "THE CENTRAL LINE WAS PLACED IN THE OR BY ANESTHESIA ON 3/8/24. WHILE TRYING TO DRAW LABS ON 3/11/24, THE RN NO TICED THAT WHILE TRYING TO PULL BACK FROM THE INTRODUCER PORT, ALL SHE COULD DRAW BACK WAS AIR. THE LINE WAS INSPECTED AND USING A 10 ML SYRINGE APPROXIMATELY 100 ML OF AIR WAS REMOVED. ALL OF THE OTHER LUMENS PULLED BACK BLOOD WITH NO ISSUE. THE PATIENT HAD NO SUBCUTANEOUS AIR PALPABLE".
Patient problems: Insufficient Information
MAUDE report 19948269 ↗InjuryFeb 8, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
IT WAS REPORTED THAT: PATIENT CODED TWICE AND HAD A CHLORHEXIDINE ALLERGIC REACTION. ADDITIONAL INFORMATION: IT WAS KNOWN THAT THE PATIENT HAD A CHG ALLERGY PRIOR TO THE PROCEDURE. HE HAD A PREVIOUS ARREST DURING A CABG BECAUSE OF THE ALLERGY. PRECAUTIONS WERE TAKEN WITH MEDICATION SELECTION AND AVOIDING ANY CHG SCRUBS OR CLEANERS FOR THE SURGERY. WHEN THE SHEATH WAS INSERTED THE PATIENT BECAME TACHYCARDIC AND HYPOTENSIVE. ONCE THE REACTION STARTED, HE WAS TREATED WITH BOLUS OF EPINEPHRINE AND A GTT WAS STARTED ALONG WITH VASOPRESSIN, COMPRESSIONS WERE STARTED FOR PERFUSION ONLY AS THE PATIENT DID NOT LOSE PULSE. ONCE THE SURGICAL TEAM REALIZED IT WAS THE SHEATH IT WAS REMOVED IMMEDIATELY AND REPLACE WITH A SHEATH THAT WAS NOT COATED. THE PATIENT WAS TRANSFERRED TO ICU FOLLOWING THE PROCEDURE WITH HEMODYNAMICS THAT WOULD BE WITHIN NORMAL LIMITS AFTER A CABG, MEANING THAT HE REQUIRED A LITTLE EXTRA BLOOD PRESSURE SUPPORT INITIALLY BUT WAS ABLE TO BE WEANED OFF. THE PATIENT DID DEVELOP AN ILEUS AFTER THE PROCEDURE BUT WAS UNLIKELY AS A RESULT OF THE ANAPHYLAXIS EVENT. THERE WAS NO FURTHER HEALTH CONSEQUENCES. THE PATIENT HAS BEEN DISCHARGED HOME WITH HOME HELP.
Patient problems: Hypersensitivity/Allergic reaction; Low Blood Pressure/ Hypotension; Tachycardia
MAUDE report 18666318 ↗InjuryAug 8, 2023Biocompatibility; Biocompatibility
MEDWATCH REPORT RECEIVED: "PATIENT (PT) HAD MULTIPLE KNOWN ALLERGIES AND UNDERWENT ALLERGY TEST FOLLOWING RECENT ANAPHYLACTIC REACTION AFTER INTUBATION. SURGERY CANCELLED AND ALLERGY TESTING DONE REVEALED CHG (CHLORHEXIDINE GLUCONATE) ALLERGY. TWO MONTHS LATER, RETURNED TO OR. AFTER CENTRAL LINE PLACED - DEVELOPED HYPOTENSION, TREATED FOR ANAPHYLAXIS. SUSPECTED SOURCE CENTRAL LINE - CHLORHEXIDINE COATING. PATIENT PREPPED WITH BETADINE FOR BOTH LINE INSERTIONS. LINE REMOVED. NEW NON-COATED CHG INSERTED. DECISION TO PROCEED WITH SURGICAL PROCEDURE. PT REQUIRED 125 MG OF SOLU-CORTEF AND HAD ALREADY REC'D DECADRON FOR PONV (POSTOPERATIVE NAUSEA AND VOMITING) PRIOR TO ANAPHYLACTIC EPISODE."
Patient problems: Anaphylactic Shock; Angioedema; Low Blood Pressure/ Hypotension; Renal Impairment; Urinary Retention
MAUDE report 17479789 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011761. One submission often covers many models, so check the product and lot details in each record.
- Z-2385-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH;
- Z-2555-2019 · Aug 6, 2019 · Terminated8.5 Fr. X 10 cm ARROW+gard Blue PSI Kit Cath-Gard for use with 7.5 - 8 Fr.Catheters Product Code: CDC-29803-1A - Product Usage: The MACTM Two-Lumen Central Venous Access Device wit
- Z-1983-2018 · Apr 11, 2018 · TerminatedMAC(TM) Two-Lumen Central Venous Access Kit with ARROWg+ard Blue(R) Access Device, Integral Hemostasis Valve and Sharps Safety Features for use with 7.5 - 8 Fr. Catheters; Product
- Z-1324-2018 · Dec 22, 2017 · Terminated(1) ARROWg+ard Blue(R) MAC(TM) Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters, Product Number: ASK-21142-RH; (2) Arrowg+ard Blue(R) MAC(TM) Two-Lumen Centra
- Z-1002-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of diseases or conditions requiring central venous
- Z-1000-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue MAC" Two-Lumen Central Venous Access Kit Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of disease
- Z-2333-2017 · May 9, 2017 · Terminated1) ARROWg+ard Blue MAC Two-Lumen Central Venous Access Kit 2) ARROWg+ard Blue PSI Kit 3) ARROWg+ard Blue MAC Two-Lumen Central Venous Access Kit 4) ARROWg+ard Blue PSI Kit The MACT
- Z-2825-2015 · Aug 27, 2015 · TerminatedArrow Kits with Medtronic Covidien 0.9% Sodium Chloride Flush Syringes
- Z-0066-2015 · Aug 6, 2014 · TerminatedArrow Select Kits (ASK) and Maximal Barrier Precautions Kits (CDC) with BD Medical Posiflush 10mL Saline Flush Syringe.
- Z-0783-2010 · Nov 24, 2009 · TerminatedMAC (TM) Two Lumen Central Venous Access Kit with ARROWg+ard Blue (R) Access Device, Integral Hemostasis Valve, Sharps Safety Features and Maximal Barrier Precautions for use with
Other Catheter, intravascular, therapeutic, short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page