ARROW IPN917233Recalls & reported problems
Teleflex Incorporated · Berman Angiographic Balloon Catheter · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902180763), received between Nov 15, 2022 and Nov 15, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K892530); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2022Leak/Splash; Leak/Splash
THE REPORT STATES THAT "AFTER 10 MINUTES OF USE, A RUPTURE OF THE CATHETER BODY OUTSIDE THE PATIENT DURING ANGIOGRAPHIC INJECTION WITH THE USUAL VOLUME AND FLOW OCCURRED. AS A RESULT, ANOTHER CATHETER WAS INSERTED IN THE VEIN JUGULAR LEFT". ADDITIONAL INFORMATION STATES THAT THE SECOND CATHETER WAS INSERTED AT THE SAME INSERTION SITE. NO MEDICAL INTERVENTION REQUIRED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15797275 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K892530. One submission often covers many models, so check the product and lot details in each record.
- Z-0715-2020 · Nov 15, 2019 · TerminatedArrow Berman Angiographic Catheter 50cc Item Number: AI-07130
- Z-0716-2020 · Nov 15, 2019 · TerminatedArrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J
- Z-1057-2013 · Mar 4, 2013 · TerminatedArrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arrow Berman(TM) Angiographic Catheter is indicated for use in ri
Other CATHETER, FLOW DIRECTED devices
- Graftmaster 1012580-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED30 deaths225 reports
- SWAN-GANZ CCOMBO V 777F8Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls92 reports
- Graftmaster 1012581-16Abbott Vascular Inc. · CATHETER, FLOW DIRECTED21 deaths147 reports
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- Graftmaster 1012581-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED12 deaths135 reports
- Graftmaster 1012582-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths77 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- Graftmaster 1012580-26Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths72 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page