ARROW IPN920005Recalls & reported problems

Teleflex Incorporated · Femoral Arterial Line Catheterization · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902192063), received between Sep 17, 2024 and Sep 17, 2024 — classified as 1 malfunction.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K810962); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 21, 2026Open, Classified

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    Z-2391-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024

    WHILE INSERTING A CENTRAL VENOUS CATHETER INTO RIGHT FEMORAL SITE, I WAS ALSO TRYING TO INSERT AN ARTERIAL LINE. AFTER VERIFYING THAT MY INTRODUCER NEEDLE WAS IN THE FEMORAL ARTERY PER US (ULTRASOUND) PLUS CLINICAL VERIFICATION PER PULSATING BLOOD RETURN FROM THE ARTERIAL NEEDLE, I INSERTED THE GUIDEWIRE THRU THE NEEDLE. THE GUIDE WIRE WENT IN SMOOTHLY. HOWEVER, WHEN I WAS RETRACTING THE NEEDLE OUT, I ENCOUNTERED RESISTANCE. I MANIPULATED THE GUIDEWIRE (MOVING IT IN AND OUT TO SEE IF THE RESISTANCE WILL DISAPPEAR). THE RESISTANCE IMPROVED; HOWEVER, WHEN I TOTALLY PULLED OUT THE NEEDLE, THE GUIDEWIRE "FRAYED". SINCE THERE WAS RESISTANCE, AS IF THE TIP OF THE GUIDEWIRE IS ANCHORED INSIDE THE SKIN, I DID NOT TRY ANY FURTHER ATTEMPT TO RECOVER THE TIP OF THE GUIDEWIRE. PATIENT WENT TO NEURO IR WHERE THE PATIENT UNDERWENT PIPELINE ASSISTED COIL EMBOLIZATION, ALSO SUCCESSFUL RETRIEVAL OF THE REMNANT OF THE GUIDEWIRE FROM THE RIGHT FEMORAL SITE. THERE WAS NO HEMATOMA, NO VASCULAR COMPROMISE OF THE LEG POST-PROCEDURE.

    Patient problems: Foreign Body In Patient

    MAUDE report 20235897 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K810962. One submission often covers many models, so check the product and lot details in each record.

  • Z-2377-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE
  • Z-1960-2018 · Apr 11, 2018 · Terminated
    Spring Wire Guide Introducer Needle; Product Codes: AN-04320 and AN-04418

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →