Skip to content

ARROW IPN920501Recalls & reported problems

Teleflex Incorporated · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10801902195422), received between Jul 30, 2026 and Sep 16, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Device Damaged Prior to Use” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged Prior to Use3
  • Material Twisted/Bent3
  • Incorrect Measurement1
  • Material Separation1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Device Damaged Prior to Use; Material Twisted/Bent

    It was reported that during inspection of the arterial catheter set, the guidewire was observed to be bent. A new arterial catheter set was obtained and the procedure was completed without reported complications. The issue was identified prior to patient use, and the device was not used. There was no patient involvement and no reports of patient injury, adverse health consequences, or clinically significant harm associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782606 ↗
  • MalfunctionSep 16, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782556 ↗
  • MalfunctionSep 16, 2026Device Damaged Prior to Use; Material Twisted/Bent

    It was reported that after opening the arterial catheter packaging, the guidewire was observed to be folded/bent. The issue was detected before patient use, and the device was not used. There was no patient involvement and no patient injury, adverse health consequences, or clinically significant harm associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782530 ↗
  • MalfunctionSep 3, 2026Unraveled Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50577504 ↗
  • InjuryJul 30, 2026Incorrect Measurement; Material Separation

    Patient problems: Foreign Body In Patient

    MAUDE report 50090928 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page