Skip to content

ARROW IPN920735Recalls & reported problems

Teleflex Incorporated · Arterial Catheterization Set · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10801902197327), received between Jul 29, 2026 and Sep 3, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Device Damaged Prior to Use” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 6 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged Prior to Use6
  • Material Twisted/Bent6
  • Failure to Obtain Sample1
  • Insufficient Device Problem Information1
  • Obstruction of Flow1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Hematoma1
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50577260 ↗
  • InjuryAug 31, 2026Insufficient Device Problem Information; Physical Resistance/Sticking

    Patient problems: Hematoma; Insufficient Information

    MAUDE report 50534276 ↗
  • MalfunctionJul 30, 2026Obstruction of Flow; Failure to Obtain Sample

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50083155 ↗
  • MalfunctionJul 29, 2026Device Damaged Prior to Use; Material Twisted/Bent

    It was reported that five arterial kits from lot number 71F26C0995 were opened for use during the same procedure. Upon opening, the spring wire guide (SWG) contained within each kit was observed to be bent/kinked. The issue was identified prior to patient use, and none of the affected devices were used. There was no patient involvement associated with this event. No patient injury, adverse health consequences, or clinically significant harm were reported. Associated MDRs include 3006425876-2026-00820 (Device #1), 3006425876-2026-00839 (Device #2), 3006425876-2026-00840 (Device #3), 3006425876-2026-00841 (Device #4)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077255 ↗
  • MalfunctionJul 29, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077244 ↗
  • MalfunctionJul 29, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077220 ↗
  • MalfunctionJul 29, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077207 ↗
  • MalfunctionJul 29, 2026Device Damaged Prior to Use; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077098 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page