ARROW IPN927347Recalls & reported problems
Teleflex Incorporated · Arrowg+ard Blue(R) Two-Lumen Hemodialysis Catheterization for High Volume Infusions · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10801902213553), received between Jan 23, 2024 and Dec 1, 2025 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K993933); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2388-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation4
- Material Too Rigid or Stiff2
- Material Twisted/Bent2
- Difficult to Advance1
- Fluid/Blood Leak1
- Material Too Soft/Flexible1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
- Laceration(s)1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT: "WHEN USING THE RAULERSON SYRINGE THE PHYSICIAN STATED THAT WHEN PULLING BACK ON THE PLUNGER NO BLOOD WAS PULLED BACK, ONLY AIR. PHYSICIAN DECIDED TO PULL ANOTHER KIT AND PLACE LINE ON THE OTHER SIDE OF THE PATIENT. THERE WAS NO REPORTED PATIENT HARM."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23680790 ↗MalfunctionMay 9, 2024Material Twisted/Bent; Material Too Soft/Flexible; Material Twisted/Bent; Material Too Soft/Flexible
IT WAS REPORTED "CATHETER WAS BEING PLACED IN THE IJ AND CUSTOMER EXPERIENCED DIFFICULTY PASSING THE CATHETER OVER THE WIRE. CUSTOMER IS REPORTING THE WIRE WAS FLIMSY AND BENT." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THERE WAS NO DELAY TO THERAPY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 19279732 ↗MalfunctionMay 9, 2024Difficult to Advance; Difficult to Advance
IT WAS REPORTED "PHYSICIAN STATED THAT CATHETER WOULD NOT PASS INTO PATIENT WIRE AND DILATOR AND WIRE WENT SMOOTH BUT CATHETER GOT HUNG UP AROUND THE SIDE HOLES CATHETER WOULDN'T GO." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. NO DELAY TO THERAPY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19278561 ↗InjuryFeb 16, 2024Material Too Rigid or Stiff; Material Too Rigid or Stiff
THE CLIENT COMPLAINED THAT THE PLASTIC ON THE CATHETER IS TOO RIDGED. THEY REPORTED THAT THIS CAUSE CAUSED BRACHIOCEPHALIC INJURY, WHICH HAD TO BE "FIXED". ADDITIONAL INFORMATION WAS REQUESTED FROM THE ACCOUNT.
Patient problems: Laceration(s)
MAUDE report 18720975 ↗InjuryFeb 16, 2024Material Too Rigid or Stiff; Material Too Rigid or Stiff
THE CLIENT COMPLAINED THAT THE PLASTIC ON THE CATHETER IS TOO RIDGED. THEY REPORTED THAT THIS CAUSE CAUSED BRACHIOCEPHALIC INJURY, WHICH HAD TO "FIXED". ADDITIONAL INFORMATION WAS REQUESTED FROM THE ACCOUNT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18719410 ↗MalfunctionJan 23, 2024Material Twisted/Bent; Material Twisted/Bent
WE ELECTED TO PLACE A TEMPORARY HD CATHETER IN THE FEMORAL VEIN ON THE R BECAUSE OF ACCESS ISSUES. WE WERE ABLE TO EASILY ACCESS THE VEIN USING US GUIDANCE. WE THREAD THE WIRE AND CONFIRMED VENOUS PLACEMENT. A LARGE SKIN INCISION WAS MADE TO ACCOMMODATE THE DILATORS AND CATHETER. THE PROBLEM BEGAN UPON DILATION. WE ENCOUNTERED MORE RESISTANCE THAN EXPECTED DURING DILATION AND UPON TRYING TO ADVANCE THE CATHETER ON THE FIRST ATTEMPT. I ATTEMPTED TO ADVANCE THE CATHETER MYSELF BUT WAS UNSUCCESSFUL. AT THAT TIME, I ELECTED TO PULL THE CATHETER OUT, CHANGE TRY ANOTHER KIT. I THREADED A CATHETER OVER THE WIRE, REMOVED THE OLD WIRE WHICH HAD A VERY SMALL KINK AND REPLACE IT WITH A NEW WIRE. I THEN RE-DILATED WITH BOTH DILATORS AND THEN TRIED TO ADVANCE THE NEW 20CM 14F HD CATHETER. I WAS UNSUCCESSFUL. I MET A SIGNIFICANT AMOUNT OF RESISTANCE, APPROXIMATELY 8-10CM IN, AND DESPITE MANY MANEUVERS TO ADVANCE, I WAS UNSUCCESSFUL. THE CATHETER WAS TOO SOFT AND WOULD START TO KINK UNDER THE SKIN. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18562364 ↗MalfunctionJan 23, 2024Material Deformation; Material Deformation
WE ELECTED TO PLACE A TEMPORARY HD CATHETER IN THE FEMORAL VEIN ON THE R BECAUSE OF ACCESS ISSUES. WE WERE ABLE TO EASILY ACCESS THE VEIN USING US GUIDANCE. WE THREAD THE WIRE AND CONFIRMED VENOUS PLACEMENT. A LARGE SKIN INCISION WAS MADE TO ACCOMMODATE THE DILATORS AND CATHETER. THE PROBLEM BEGAN UPON DILATION. WE ENCOUNTERED MORE RESISTANCE THAN EXPECTED DURING DILATION AND UPON TRYING TO ADVANCE THE CATHETER ON THE FIRST ATTEMPT. I ATTEMPTED TO ADVANCE THE CATHETER MYSELF BUT WAS UNSUCCESSFUL. AT THAT TIME, I ELECTED TO PULL THE CATHETER OUT, CHANGE TRY ANOTHER KIT. I THREADED A CATHETER OVER THE WIRE, REMOVED THE OLD WIRE WHICH HAD A VERY SMALL KINK AND REPLACE IT WITH A NEW WIRE. I THEN RE-DILATED WITH BOTH DILATORS AND THEN TRIED TO ADVANCE THE NEW 20CM 14F HD CATHETER. I WAS UNSUCCESSFUL. I MET A SIGNIFICANT AMOUNT OF RESISTANCE, APPROXIMATELY 8-10CM IN, AND DESPITE MANY MANEUVERS TO ADVANCE, I WAS UNSUCCESSFUL. THE CATHETER WAS TOO SOFT AND WOULD START TO KINK UNDER THE SKIN. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18562015 ↗MalfunctionJan 23, 2024Material Deformation; Material Deformation
WE ELECTED TO PLACE A TEMPORARY HD CATHETER IN THE FEMORAL VEIN ON THE R BECAUSE OF ACCESS ISSUES. WE WERE ABLE TO EASILY ACCESS THE VEIN USING US GUIDANCE. WE THREAD THE WIRE AND CONFIRMED VENOUS PLACEMENT. A LARGE SKIN INCISION WAS MADE TO ACCOMMODATE THE DILATORS AND CATHETER. THE PROBLEM BEGAN UPON DILATION. WE ENCOUNTERED MORE RESISTANCE THAN EXPECTED DURING DILATION AND UPON TRYING TO ADVANCE THE CATHETER ON THE FIRST ATTEMPT. I ATTEMPTED TO ADVANCE THE CATHETER MYSELF BUT WAS UNSUCCESSFUL. AT THAT TIME, I ELECTED TO PULL THE CATHETER OUT, CHANGE TRY ANOTHER KIT. I THREADED A CATHETER OVER THE WIRE, REMOVED THE OLD WIRE WHICH HAD A VERY SMALL KINK AND REPLACE IT WITH A NEW WIRE. I THEN RE-DILATED WITH BOTH DILATORS AND THEN TRIED TO ADVANCE THE NEW 20CM 14F HD CATHETER. I WAS UNSUCCESSFUL. I MET A SIGNIFICANT AMOUNT OF RESISTANCE, APPROXIMATELY 8-10CM IN, AND DESPITE MANY MANEUVERS TO ADVANCE, I WAS UNSUCCESSFUL. THE CATHETER WAS TOO SOFT AND WOULD START TO KINK UNDER THE SKIN. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18561999 ↗MalfunctionJan 23, 2024Material Deformation; Material Deformation
WE ELECTED TO PLACE A TEMPORARY HD CATHETER IN THE FEMORAL VEIN ON THE R BECAUSE OF ACCESS ISSUES. WE WERE ABLE TO EASILY ACCESS THE VEIN USING US GUIDANCE. WE THREAD THE WIRE AND CONFIRMED VENOUS PLACEMENT. A LARGE SKIN INCISION WAS MADE TO ACCOMMODATE THE DILATORS AND CATHETER. THE PROBLEM BEGAN UPON DILATION. WE ENCOUNTERED MORE RESISTANCE THAN EXPECTED DURING DILATION AND UPON TRYING TO ADVANCE THE CATHETER ON THE FIRST ATTEMPT. I ATTEMPTED TO ADVANCE THE CATHETER MYSELF BUT WAS UNSUCCESSFUL. AT THAT TIME, I ELECTED TO PULL THE CATHETER OUT, CHANGE TRY ANOTHER KIT. I THREADED A CATHETER OVER THE WIRE, REMOVED THE OLD WIRE WHICH HAD A VERY SMALL KINK AND REPLACE IT WITH A NEW WIRE. I THEN RE-DILATED WITH BOTH DILATORS AND THEN TRIED TO ADVANCE THE NEW 20CM 14F HD CATHETER. I WAS UNSUCCESSFUL. I MET A SIGNIFICANT AMOUNT OF RESISTANCE, APPROXIMATELY 8-10CM IN, AND DESPITE MANY MANEUVERS TO ADVANCE, I WAS UNSUCCESSFUL. THE CATHETER WAS TOO SOFT AND WOULD START TO KINK UNDER THE SKIN. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18561983 ↗MalfunctionJan 23, 2024Material Deformation; Material Deformation
WE ELECTED TO PLACE A TEMPORARY HD CATHETER IN THE FEMORAL VEIN ON THE R BECAUSE OF ACCESS ISSUES. WE WERE ABLE TO EASILY ACCESS THE VEIN USING US GUIDANCE. WE THREAD THE WIRE AND CONFIRMED VENOUS PLACEMENT. A LARGE SKIN INCISION WAS MADE TO ACCOMMODATE THE DILATORS AND CATHETER. THE PROBLEM BEGAN UPON DILATION. WE ENCOUNTERED MORE RESISTANCE THAN EXPECTED DURING DILATION AND UPON TRYING TO ADVANCE THE CATHETER ON THE FIRST ATTEMPT. I ATTEMPTED TO ADVANCE THE CATHETER MYSELF BUT WAS UNSUCCESSFUL. AT THAT TIME, I ELECTED TO PULL THE CATHETER OUT, CHANGE TRY ANOTHER KIT. I THREADED A CATHETER OVER THE WIRE, REMOVED THE OLD WIRE WHICH HAD A VERY SMALL KINK AND REPLACE IT WITH A NEW WIRE. I THEN RE-DILATED WITH BOTH DILATORS AND THEN TRIED TO ADVANCE THE NEW 20CM 14F HD CATHETER. I WAS UNSUCCESSFUL. I MET A SIGNIFICANT AMOUNT OF RESISTANCE, APPROXIMATELY 8-10CM IN, AND DESPITE MANY MANEUVERS TO ADVANCE, I WAS UNSUCCESSFUL. THE CATHETER WAS TOO SOFT AND WOULD START TO KINK UNDER THE SKIN. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18561800 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K993933. One submission often covers many models, so check the product and lot details in each record.
- Z-2373-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYS
- Z-1336-2018 · Dec 22, 2017 · TerminatedARROWg+ard Blue Two-Lumen Hemodialysis Catheterization Kit for High Volume Infusions (1) Part Number (PN): CDC-22122-1A; (2) PN: CDC-25122-1A; (3) PN: CDC-25142-1A; (4) CDC-26142-1
- Z-0658-2018 · Oct 17, 2017 · TerminatedARROWg+ard Blue Two Lumen Hemodialysis Catheterization Kit for High Volume Infusions; Product Codes: AK-22142-CDC, AK-22142-F, AK-25142-CDC, AK-25142-F; Exp. Dates Feb 2018 - Nov 2
- Z-2334-2017 · May 9, 2017 · Terminated1) ARROWg+ard Blue You-Blend Two-Lumen Hemodialysis Catheterization Kit for High Volume Infusions The large-bore two-lumen catheter permits venous access to the central circulation
- Z-0639-04 · Feb 18, 2004 · TerminatedTwo Lumen 14F Hemodialysis Catheter Kits/Sets with Blue FlexTip Arrowg+ard Blue Curved Catheter for high Volume Ifusions
Other Catheter, hemodialysis, non-implanted devices
- MAHURKAR 8888135191Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall175 reports
- MAHURKAR 8888135241Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall1 death100 reports
- ARROW IPN036723Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls2 reports
- DuoGlide catheter 5625240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- ARROW IPN036412Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- ARROW IPN036419Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN927343Teleflex Incorporated · Catheter, hemodialysis, non-implanted2 recalls0 reports
- ARROW IPN036428Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page