Skip to content

ARROW IPN929209Recalls & reported problems

Teleflex Incorporated · FiberOptix(TM) IAB · Class 2

Quick summary

This Teleflex Arrow FiberOptix intra-aortic balloon is a balloon placed in the aorta to support the heart, used with a control console. Many of the 112 reports describe leaks, false alarms, and inflation problems, and some describe a balloon rupture, a cracked connector, or a kinked helium line. The reports include 3 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
121
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 121 adverse event reports that list this exact device (DI 10801902218916), received between Feb 7, 2025 and Sep 29, 2026 — classified as 3 injuries, 118 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (25 reports).

1 other recall cover devices cleared under the same FDA submission (K200634); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 3 reportsApr 2025: 2 reportsMay 2025: 13 reportsJun 2025: 10 reportsJul 2025: 2 reportsAug 2025: 8 reportsSep 2025: 4 reportsOct 2025: 7 reportsNov 2025: 10 reportsDec 2025: 14 reportsJan 2026: 5 reportsFeb 2026: 6 reportsMar 2026: 8 reportsApr 2026: 12 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 10 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash25
  • Device Alarm System21
  • Fluid/Blood Leak15
  • Inflation Problem14
  • Material Integrity Problem9
  • Connection Problem7
  • Material Rupture6
  • Patient Device Interaction Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions95
  • Insufficient Information26

Recent adverse event reports

The 25 most recent of 121 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Inflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933054 ↗
  • MalfunctionSep 17, 2026Gas/Air Leak

    Patient problems: Insufficient Information

    MAUDE report 50809915 ↗
  • MalfunctionAug 31, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50531192 ↗
  • MalfunctionAug 31, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50531114 ↗
  • MalfunctionAug 24, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50410590 ↗
  • MalfunctionAug 24, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50409956 ↗
  • MalfunctionAug 17, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50286957 ↗
  • MalfunctionAug 11, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50202451 ↗
  • MalfunctionAug 10, 2026Fluid/Blood Leak

    It was reported that threads stripped on actis extraction tip.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50198184 ↗
  • MalfunctionAug 10, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50194695 ↗
  • MalfunctionAug 10, 2026Device Handling Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50194638 ↗
  • MalfunctionAug 3, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50118642 ↗
  • MalfunctionMay 18, 2026Material Twisted/Bent

    IT WAS REPORTED THAT "CALLER REPORTS POSSIBLE HE LOSS 3 AND HIGH BASELINE ALARMS. HELIUM DRIVELINE FOUND TO BE SEVERELY KINKED. CALLER HAS RETRIEVED ANOTHER HELIUM DRIVELINE. HELIUM DRIVELINE EXCHANGED SUCCESSFULLY. HE ALARMS RESOLVED AFTER EXCHANGE". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25206072 ↗
  • MalfunctionApr 23, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT BLOOD WAS DISCOVERED IN THE HELIUM DRIVELINE. AS A RESULT THE IAB WAS REMOVED. IAB WAS NOT REPLACED. NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24971641 ↗
  • MalfunctionApr 22, 2026Material Rupture

    REPORTED AS "BALLOON RUPTURE". AS A RESULT, THE IAB WAS REMOVED. THE IAB WAS NOT REPLACED. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956680 ↗
  • MalfunctionApr 22, 2026Leak/Splash

    REPORTED AS "POTENTIAL IAB RUPTURE". THE REPORT FURTHER STATES, "[USER] IS CALLING FROM THE CCL. THEY HAVE JUST PLACED A 40 CC LWS AND ARE GETTING POSSIBLE HE LOSS 2 ALARMS. THEY HAVE RECEIVED 5 TOTAL AND THEY ARE COMING APPROXIMATELY EVERY 2 MINS. HE IS REQUESTING ASSISTANCE. [USER] CONFIRMED THERE WAS NO BLOOD IN DRIVELINE, NO VISIBLE KINKING, AND CONNECTIONS WERE TIGHT. CCL TEAM OPTING TO REPLACE IAB. MD REMOVED IAB AND SHEATH OVER WIRE PRESERVING THE L FEM ACCESS FOR NEW 40 CC IAB REPLACEMENT. IAB REPLACED AND IABP THERAPY STARTED WITHOUT ALARMS". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24955379 ↗
  • MalfunctionApr 20, 2026Crack; Disconnection

    IT WAS REPORTED THAT "LUER LOCK CONNECTOR CRACKED AND TUBING DISCONNECTED - CRACKED CONNECTOR SEQUESTERED; AFTER TURN, RN NOTICED BALLOON HAD PULLED OUT FURTHER FROM THE PROTECTIVE COVERING". ADDITIONAL INFORMATION RECEIVED STATES THAT, "THE ACCOUNT IS NOT SHARING ANY ADDITIONAL INFORMATION OTHER THAN WHAT WAS PROVIDED".

    Patient problems: Insufficient Information

    MAUDE report 24927627 ↗
  • MalfunctionApr 20, 2026Material Rupture

    IT WAS REPORTED THAT THERE WAS A BALLOON RUPTURE. THE INFORMATION FURTHER STATES, "I SPOKE WITH THE CLINICAL ICU EDUCATOR WHO LET ME KNOW THAT THE PATIENT WAS KNOWN TO HAVE TORTUOUS ANATOMY WHICH MAY HAVE CONTRIBUTED TO THE RUPTURE. BALLOON REMOVED, PER ACCOUNT THERE MAY HAVE BEEN A DELAY IN BALLOON REMOVAL. UNKNOWN AT THIS TIME HOW MUCH OF A DELAY". NO REPORT OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926440 ↗
  • MalfunctionApr 13, 2026Material Integrity Problem

    IT WAS REPORTED "BALLOON FAILED TO UNWRAP". ADDITIONAL INFOMORMATION STATES THAT THE BALLOON CATHETER "EVENTUALLY UNWRAPPED". THE IAB CATHETER WAS NOT REMOVED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "CRITICAL".

    Patient problems: Insufficient Information

    MAUDE report 24860176 ↗
  • MalfunctionApr 9, 2026Inflation Problem

    IT WAS REPORTED "BALLOON DID NOT UNWRAP COMPLETELY AND NEEDED TO BE REPLACED". ADDITIONAL INFORMATION STATES THAT MANUAL INFLATION WAS USED TO CONTINUE THERAPY. THE ORIGIONAL CATHETER REMAINED IN PLACE AND WAS ADEQUATE FOR THERAPY. NO PAITNE INJURY OR CONSEQUENCE REPORTED. THE PATIENTS CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Insufficient Information

    MAUDE report 24829203 ↗
  • MalfunctionApr 8, 2026Inflation Problem

    IT WAS REPORTED "BALLOON DID NOT UNWRAP COMPLETELY AND NEEDED TO BE REPLACED". ADDITIONAL INFORMATION STATES THAT LOCATION OF THE SECOND CATHETER USED IS UNKNOWN. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817048 ↗
  • MalfunctionApr 8, 2026Inflation Problem

    IT WAS REPORTED "BALLOON [CATHETER]HAVE NOT UNRAVELED COMPLETELY ONCE THEY HAVE BEEN PLACED IN THE BODY". ADDITIONAL INFORMATION STATES THAT MANUAL INFLATION WAS USED AND THE BALLOON INFLATED AND WAS ADEQUATE FOR THERAPY. NO PATIENT INJURY OR COSNEQUENCE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24816855 ↗
  • MalfunctionApr 7, 2026Inflation Problem

    IT WAS REPORTED " BALLOON DID NOT INFLATE". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24800258 ↗
  • MalfunctionApr 6, 2026Material Rupture

    IT WAS REPORTED "BLOOD IN HE DRIVELINE, IAB RUPTURE". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788972 ↗
  • MalfunctionApr 1, 2026Device Sensing Problem

    IT WAS REPORTED "AP FOS OUT OF RANGE". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR COSNEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24755375 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K200634. One submission often covers many models, so check the product and lot details in each record.

  • Z-1878-2024 · Apr 29, 2024 · Open, Classified
    Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS

Other System, balloon, intra-aortic and control devices

All System, balloon, intra-aortic and control devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.